FDA Adverse Event Malfunction Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA

MDR report key: 1031380 · Received April 18, 2008

Report

Report Number
2023826-2008-00552
Event Type
Malfunction
Date Received
April 18, 2008
Date of Event
November 19, 2007
Report Date
March 24, 2008
Manufacturer
STAAR SURGICAL
Product Code
HQL
PMA / PMN Number
P990013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

RESULTS: A VISUAL INSPECTION OF THE RETURNED PRODUCT SHOWS THE LENS IS STUCK IN THE TIP OF THE CARTRIDGE AND THE CARTRIDGE IS DAMAGED. THE INJECTOR HAS NO VISIBLE DAMAGE. THE CARTRIDGE AND INJECTOR HAS CLEAR SURGICAL RESIDUE ON THEM. CONCLUSIONS: A MULTIFUNCTIONAL TEAM INVESTIGATED COMPLAINTS REGARDING LENS TEARS ASSOCIATED WITH THE CQ CARTRIDGE: THE RESULTANT CORRECTIVE ACTIONS INCLUDED IMPROVEMENTS IN MFG, RELEASE TESTING, AND EVAL OF TEST RESULTS. ADDITIONALLY, THE INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS.

Description of Event or Problem · 1

THE REPORTER STATED THAT THE SURGEON WAS ADVANCING A THREE PIECE COLLAMER LENS MODEL CQ2015 AND THE LENS GOT STUCK IN THE CARTRIDGE. NO PT CONTACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE FOLDABLE INTRAOCULA INTRAOCULAR LENS HQL STAAR SURGICAL CQ2015 NA

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE MODEL CQ CARTRIDGE-FP| INJECTOR MODEL MSI-PM