FDA Adverse Event Malfunction Summary report: N

PUMP MMT-1780KPK 670G PATHWAY BLACK MG

MDR report key: 10313229 · Received July 23, 2020

Report

Report Number
2032227-2020-151348
Event Type
Malfunction
Date Received
July 23, 2020
Date of Event
July 17, 2020
Report Date
May 3, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
00763000090203
Removal / Correction Number
Z-0956-2020
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. UPDATED H9: Z-0956-2020. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE WAS RECEIVED WITH A CRACKED CASE CORNER OF THE BELT CLIP RAILS NEAR THE BATTERY COMPARTMENT. THE TEST RESERVOIR DOES LOCK PROPERLY INSIDE THE RESERVOIR TUBE. HOWEVER, DEVICE WAS ALSO RECEIVED WITH A CRITICAL PUMP ERROR ALARM AND PUMP ERROR 35 ALARM DUE TO MOISTURE DAMAGE ON THE FORCE SENSOR. UNABLE TO PERFORM THE SELF TEST, DISPLACEMENT TEST, REWIND TEST, PRIME OR SEATING TEST, BASIC OCCLUSION TEST, FORCE SENSOR TEST, OCCLUSION TEST, SLEEP CURRENT MEASUREMENT TEST, AND ACTIVE CURRENT MEASUREMENT TEST DUE TO CRITICAL PUMP ERROR ALARM. MOISTURE DAMAGE WAS ALSO FOUND ON THE ELECTRONIC ASSEMBLY, MOTOR AND VIBRATOR DURING VISUAL INSPECTION.

Description of Event or Problem · 1

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP ALARMED BROKEN FORCE SENSOR WAS DETECTED DURING SEATING OR REGULAR DELIVERY ERROR AND CRITICAL PUMP ERROR. THE INSULIN PUMP WAS EXPOSED TO RIVER WATER WHILE HE WENT TO RIVER RAFT RIDE. THE CUSTOMER ALSO STATED THAT THERE WAS CRACK AT BACK OF INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
777115 PUMP MMT-1780KPK 670G PATHWAY BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1780KPK HG2GQGS 00763000090203

Patients

Seq Age Sex Outcome Treatment
1 36 YR Male