FDA Adverse Event Injury Summary report: N

RAPIDLAB 1245 BLOOD GAS SYSTEM

MDR report key: 1031276 · Received April 18, 2008

Report

Report Number
1217157-2008-00003
Event Type
Injury
Date Received
April 18, 2008
Date of Event
March 27, 2008
Report Date
April 15, 2008
Manufacturer
SIEMENS MEDICAL SOLUTIONS DIAGNOSTICS, LTD
Product Code
CHL
PMA / PMN Number
K031560
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

METHOD AND RESULTS: (ACTUAL DEVICE INVOLVED IN INCIDENT WAS EVALUATED): CUSTOMER REPORTED, THE DRS VIEWING THE RESULTS IN PROGRESS DID NOT REALIZE THAT THE NUMBERS IN YELLOW BOXES ARE AN INDICATOR THAT THE SYSTEM IS STILL MEASURING THE BLOOD. THE SYSTEM DID PROVIDE THE CORRECT RESULT, IT WAS CORRECTLY DISPLAYED ON BOTH THE SCREEN AND ON THE ROLL PRINTER. CONCLUSIONS: CUSTOMER FURTHER REPORTS THE SYSTEM DID PROVIDE THE CORRECT RESULT. CUSTOMER STATED THE CORRECT RESULT WAS CORRECTLY DISPLAYED ON BOTH THE SCREEN AND ON THE ROLL PRINTER.

Description of Event or Problem · 1

CUSTOMER REPORTED, THEY HAD ONE BLOOD SAMPLE WITH A PH OF 7.00 ON THE INSTRUMENT DISPLAY AND 7.3X ON THE INSTRUMENT PRINT OUT. CUSTOMER REPORTED THAT FOR THIS REASON, A C-SECTION (ABDOMINAL DELIVERY) WAS BEGUN ON THE PT. THE CUSTOMER FURTHER REPORTED THAT WHEN THE C-SECTION WAS COMPLETED AND WHEN THEY CAME BACK TO THE INSTRUMENT, THE PRINTOUT WAS 7.3 AND IT WAS DETERMINED THAT THE C-SECTION WAS NOT NECESSARY. RESULTS OBTAINED BY PHYSICIAN WERE READ FROM INSTRUMENT DISPLAY BEFORE THE TEST WAS COMPLETED. THE CORRECT FINAL READING FOR PH WOULD HAVE BEEN DISPLAYED WITHIN 30 SECS. THE CORRECT RESULT WAS DISPLAYED IN BLACK NUMBERS ON A BLUE BACKGROUND. THE PHYSICIAN PREMATURELY ACTED ON A RESULT POSTED IN A YELLOW BOX ON DISPLAY INDICATING RESULTS UPDATING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RAPIDLAB 1245 BLOOD GAS SYSTEM RAPIDLAB 1245 CHL SIEMENS MEDICAL SOLUTIONS DIAGNOSTICS, LTD RL1245 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK