SYRINGE 20ML LL 120/PKG
Report
- Report Number
- 3003152976-2020-00315
- Event Type
- Malfunction
- Date Received
- July 23, 2020
- Date of Event
- July 2, 2020
- Report Date
- August 20, 2020
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
D.10. DEVICE AVAILABLE FOR EVAL?: YES. D.10. RETURNED TO MANUFACTURER ON: 7/27/2020. H.6. INVESTIGATION: ONE SAMPLE OF LOT 2002274 ALONG WITH THE SAMPLE CONNECTOR USED WITH SYRINGE, WAS RETURNED TO OUR QUALITY TEAM FOR INVESTIGATION. THE SAMPLES WERE VISUALLY INSPECTED, THERE WAS NO DAMAGE TO THE CONNECTION THREAD OF THE PRODUCT. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 2002274, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES, INCLUDING TIP AND THREAD VERIFICATIONS. TESTING WAS PERFORMED ON THE PRODUCT RETURNED AND FOUND THE LUER DID COMPLY WITH ISO STANDARDS. FUNCTIONAL TESTING WAS COMPLETED, FILLING THE SYRINGE WITH WATER AND CONNECTING TO THE CONNECTION DEVICE THAT WAS USED. WATER PASSED THROUGH DEVICE AND NO LEAK WAS OBSERVED. IT WAS NOTED ONE OF THE CONNECTION DEVICES, WHICH IS NOT MANUFACTURED BY BD, APPEARED TO HAVE BEEN OVER SCREWED AND THE CONNECTION STICK WAS MOVING, WHICH MAY HAVE CONTRIBUTED TO THE LEAK REPORTED. BASED ON OUR INVESTIGATION, NO DEFECT WAS OBSERVED ON OUR SYRINGE, AND DEVICE RECORDS PROVIDED SHOW NO INCIDENTS RELATED TO THE REPORTED FAILURE, THEREFORE A ROOT CAUSE RELATED TO OUR PRODUCT AND MANUFACTURING PROCESS CANNOT BE IDENTIFIED AT THIS TIME.
IT WAS REPORTED THAT SYRINGE 20ML LL 120/PKG LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE CUSTOMER HAS EXPLAINED THAT FOR THE LAST SIX WEEKS THAT WHILE PREPARING CHEMOTHERAPY IN THE ASEPTIC UNIT THAT CHEMO HAS LEAKED BETWEEN BD SYRINGES AND THE CLOSED SYSTEM TRANSFER DEVICE USED BELOW. THEY WERE WONDERING WHETHER THERE HAVE BEEN ANY MANUFACTURING OR ANY CHANGES MADE TO THE BD SYRINGES IN THE LAST WHILE. THIS HAS BEEN AN ONGOING ISSUE FOR THE LAST 6 WEEKS. THEY BELIEVE THE ISSUE RELATES TO THE CLOSED CHEMO DEVICE BUT AS THIS IS SUCH A SAFETY ISSUE WOULD LIKE TO INVESTIGATE EVERY ASPECT OF THE DRAWING UP PROCESS (INCLUDING OUR SYRINGES). THE CUSTOMER CAN MAKE AVAILABLE ANY SAMPLES WE WANT. CAN THE INVESTIGATING TEAM ADVICE WHAT THEY WOULD LIKE. CLOSED SYSTEM TRANSFER DEVICE DETAILS: BRAND NAME: CHEMOCLAVE. MANUFACTURER: ICUMEDICAL. DISTRIBUTOR (IRELAND): FANNIN (WE ARE CURRENTLY LIAISING WITH THEM CLOSELY ON THIS. DEVICE NAME: SPINNING SPIROS CLOSE MALE LUER, RED CAP. DEVICE REFERENCE CODE: CH2000S-C."
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT SYRINGE 20 ML LL 120/PKG LEAKED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "THE CUSTOMER HAS EXPLAINED THAT FOR THE LAST SIX WEEKS THAT WHILE PREPARING CHEMOTHERAPY IN THE ASEPTIC UNIT THAT CHEMO HAS LEAKED BETWEEN BD SYRINGES AND THE CLOSED SYSTEM TRANSFER DEVICE USED BELOW. THEY WERE WONDERING WHETHER THERE HAVE BEEN ANY MANUFACTURING OR ANY CHANGES MADE TO THE BD SYRINGES IN THE LAST WHILE. THIS HAS BEEN AN ONGOING ISSUE FOR THE LAST 6 WEEKS. THEY BELIEVE THE ISSUE RELATES TO THE CLOSED CHEMO DEVICE BUT AS THIS IS SUCH A SAFETY ISSUE WOULD LIKE TO INVESTIGATE EVERY ASPECT OF THE DRAWING UP PROCESS (INCLUDING OUR SYRINGES). THE CUSTOMER CAN MAKE AVAILABLE ANY SAMPLES WE WANT. CAN THE INVESTIGATING TEAM ADVICE WHAT THEY WOULD LIKE. CLOSED SYSTEM TRANSFER DEVICE DETAILS: BRAND NAME: CHEMOCLAVE. MANUFACTURER: ICUMEDICAL. DISTRIBUTOR (B)(6): FANNIN (WE ARE CURRENTLY LIAISING WITH THEM CLOSELY ON THIS. DEVICE NAME: SPINNING SPIROS CLOSE MALE LUER, RED CAP. DEVICE REFERENCE CODE: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 782405 | SYRINGE 20ML LL 120/PKG | SYRINGE | FMF | BECTON DICKINSON, S.A. | 2002274 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |