SYRINGE 0.5ML 29GA 12.7MM BLISTER 200BX
Report
- Report Number
- 1920898-2020-00920
- Event Type
- Malfunction
- Date Received
- July 22, 2020
- Date of Event
- July 12, 2020
- Report Date
- July 15, 2020
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- PHARMACIST
Narratives
INVESTIGATION SUMMARY: CUSTOMER RETURNED PHOTOS OF A 1/2CC SYRINGE. CUSTOMER STATES THAT THERE IS CONTAMINATION IN THE SYRINGE. THE ATTACHED PHOTOS WERE EXAMINED AND EXHIBITED A STRAND OF MATERIAL ON THE BARREL AND PLUNGER ROD. IT IS DIFFICULT TO DETERMINE THE IDENTITY OF THIS MATERIAL SOLELY FROM THE ATTACHED PHOTOS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 9252463. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WAS ONE (1) NOTIFICATION [200835583] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. OCCURRENCE: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 1ST RELATED COMPLAINT FOR FOREIGN MATTER IN SYRINGE ON LOT # 9140683. A REVIEW OF RISK MANAGEMENT 150RMN-0001-16 REVISION 13 INDICATES THAT THE POTENTIAL RISK OF THIS SPECIFIC REPORTED INCIDENT (SYRINGE, FOREIGN MATTER IN SYRINGE) WAS CAPTURED AND ADDRESSED APPROPRIATELY. INVESTIGATION CONCLUSION: CONFIRMED: BD WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. ROOT CAUSE DESCRIPTION: ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS IT IS DIFFICULT TO DETERMINE THE IDENTITY OF THIS MATERIAL SOLELY FROM THE ATTACHED PHOTOS. RATIONALE: BASED ON THE INVESTIGATION, NO ADDITIONAL INVESTIGATION AND NO CAPA IS REQUIRED AT THIS TIME.
IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND IN THE SYRINGE 0.5ML 29GA 12.7MM BLISTER 200BX BEFORE USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "CONTAMINATION IN THE SYRINGE. IT SEEMS TO BE A TEXTILE MATERIAL WHICH THE CUSTOMER FOUND."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 776023 | SYRINGE 0.5ML 29GA 12.7MM BLISTER 200BX | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 9140683 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |