FDA Adverse Event
Malfunction
Summary report: N
GE OEC 9800
MDR report key: 1030893
·
Received April 17, 2008
Report
- Report Number
- 1720753-2008-20637
- Event Type
- Malfunction
- Date Received
- April 17, 2008
- Date of Event
- April 2, 2008
- Report Date
- April 14, 2008
- Manufacturer
- GE OEC MEDICAL SYSTEMS INC.
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE INTERCONNECT CABLE AND THE PRINTER WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE 9800 SYSTEM HAD DAMAGED INSULATION ON THE INTERCONNECT CABLE AND THE PRINTER WOULD NOT WORK PROPERLY. NO PATIENT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GE OEC 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |