FDA Adverse Event Malfunction Summary report: N

BARDEX ALL-SILICONE FOLEY CATHETER

MDR report key: 10308711 · Received July 22, 2020

Report

Report Number
1018233-2020-04695
Event Type
Malfunction
Date Received
July 22, 2020
Report Date
October 16, 2020
Manufacturer
C.R. BARD, INC. (COVINGTON) -1018233
Product Code
EZL
UDI-DI
00801741029776
PMA / PMN Number
K040504
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE REPORTED EVENT WAS INCONCLUSIVE. NO SAMPLE WAS RETURNED FOR EVALUATION. A POTENTIAL ROOT CAUSE FOR THIS FAILURE COULD BE "HIGH MODULUS SILICONE/NON-UNIFORM THIN SAC AND /OR FINISH DIP COVERAGE ". THE LOT NUMBER WAS UNKNOWN; THEREFORE, THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED. THE INSTRUCTIONS FOR USE WERE FOUND ADEQUATE AND STATE THE FOLLOWING: "CAUTION: FEDERAL (U.S.A.) LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN. STERILE: UNLESS PACKAGE IS OPENED OR DAMAGED. WARNING: DO NOT USE OINTMENTS OR LUBRICANTS HAVING A PETROLATUM BASE. THEY WILL DAMAGE SILICONE AND MAY CAUSE BALLOON TO BURST. CAUTION: DO NOT ASPIRATE URINE THROUGH DRAINAGE FUNNEL WALL. STORAGE: STORE CATHETERS AT ROOM TEMPERATURE AWAY FROM DIRECT EXPOSURE TO LIGHT, PREFERABLY IN THE ORIGINAL BOX. SINGLE USE ONLY. DO NOT REUSE. DO NOT RESTERILIZE. FOR UROLOGICAL USE ONLY. VALVE TYPE: USE LUER SLIP SYRINGE. DO NOT USE NEEDLE. RECOMMENDED INFLATION CAPACITIES 3CC BALLOON: USE 5ML STERILE WATER 5CC BALLOON: USE 10ML STERILE WATER 30CC BALLOON: USE 35ML STERILE WATER DO NOT EXCEED RECOMMENDED CAPACITIES. WARNING: AFTER USE, THIS PRODUCT MAY BE A POTENTIAL BIOHAZARD. HANDLE AND DISPOSE OF IN ACCORDANCE WITH APPLICABLE LOCAL, STATE, AND FEDERAL LAWS AND REGULATIONS. NOTE: AGGRESSIVE TRACTION, PARTICULARLY IN THE PRESENCE OF SUTURING, IS NOT RECOMMENDED FOR 100% SILICONE FOLEY CATHETERS. BARD AND BARDEX ARE REGISTERED TRADEMARKS OF C. R. BARD, INC. OR AN AFFILIATE. TO DEFLATE CATHETER BALLOON: GENTLY INSERT A SYRINGE IN THE CATHETER VALVE. NEVER USE MORE FORCE THAN IS REQUIRED TO MAKE THE SYRINGE ¿STICK¿ IN THE VALVE. IF YOU NOTICE SLOW OR NO DEFLATION, RE-SEAT THE SYRINGE GENTLY. ALLOW THE BALLOON TO DEFLATE SLOWLY ON ITS OWN. DO NOT ASPIRATE OR MANUALLY ACCELERATE THE DEFLATION OF THE BALLOON. IF PERMITTED BY HOSPITAL PROTOCOL, THE VALVE ARM MAY BE SEVERED. IF THIS FAILS, CONTACT ADEQUATELY TRAINED PROFESSIONAL FOR ASSISTANCE, AS DIRECTED BY HOSPITAL PROTOCOL. SHOULD BALLOON RUPTURE OCCUR, CARE SHOULD BE TAKEN TO ASSURE THAT ALL BALLOON FRAGMENTS HAVE BEEN REMOVED FROM THE PATIENT. VISUALLY INSPECT THE PRODUCT FOR ANY IMPERFECTIONS OR SURFACE DETERIORATION PRIOR TO USE." H11:SECTION A THROUGH F - THE INFORMATION PROVIDED BY BD REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO BD. H3 OTHER TEXT : THE DEVICE WAS NOT RETURNED

Description of Event or Problem · 0

IT WAS REPORTED THAT THE FOLEY CATHETER WAS INSERTED IN OR AND DURING POST OPERATIVE DAY 1, THE PATIENT WAS UP FOR A WALK AND THE FOLEY FELL OUT. PER FOLEY CATHETER INSPECTION, IT WAS FOUND THAT THERE WAS A SPLIT ON THE BALLOON PART OF THE FOLEY CATHETER. ALSO IT WAS FOUND THAT THE PROBLEM WAS RECURRING. IT WAS FURTHER REPORTED THAT THE BALLOON'S PIECES APPEAR TO BE INTACT, NO MISSING PIECES.

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS STILL IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. THE DEVICE WAS NOT RETURNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE FOLEY CATHETER WAS INSERTED IN OR AND DURING POST OPERATIVE DAY 1, THE PATIENT WAS UP FOR A WALK AND THE FOLEY FELL OUT. PER FOLEY CATHETER INSPECTION, IT WAS FOUND THAT THERE WAS A SPLIT ON THE BALLOON PART OF THE FOLEY CATHETER. ALSO IT WAS FOUND THAT THE PROBLEM WAS RECURRING. IT WAS FURTHER REPORTED THAT THE BALLOON'S PIECES APPEAR TO BE INTACT, NO MISSING PIECES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
776117 BARDEX ALL-SILICONE FOLEY CATHETER FOLEY CATHETER (SILICONE) EZL C.R. BARD, INC. (COVINGTON) -1018233 165820 UNK 00801741029776

Patients

Seq Age Sex Outcome Treatment
1