SWAN-GANZ CONTINUOUS CARDIAC OUTPUT CATHETER
Report
- Report Number
- 6000002-2008-06579
- Event Type
- Malfunction
- Date Received
- April 17, 2008
- Date of Event
- November 12, 2007
- Report Date
- November 12, 2007
- Manufacturer
- EDWARDS LIFESCIENCES, PR
- Product Code
- DYG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
CATHETER #1 LEAK TESTING CONFIRMED LEAKAGE FROM THE INFLATION LUMEN DUE TO A SPLIT 0.110" LONG AT THE DISTAL EDGE OF THE THERMAL FILAMENT (MIDDLE FORM). THE TUBING SPLIT ENTERS THE INFLATION LUMEN. CATHETER #2 LEAK TESTING CONFIRMED LEAKAGE FROM THE INFLATION LUMEN DUE TO TWO SPLITS 0.012" AND .005" LONG AT THE DISTAL EDGE OF THE THERMAL FILAMENT (MIDDLE FORM). THE TUBING SPLITS ENTER THE INFLATION LUMEN. CATHETER #3 LEAK TESTING CONFIRMED LEAKAGE FROM THE INFLATION LUMEN DUE TO A SPLIT 0.090" LONG AT THE DISTAL EDGE OF THE THERMAL FILAMENT (MIDDLE FORM). THE TUBING SPLIT ENTERS THE INFLATION LUMEN. A CAPA IS IN PROCESS FOR SLIT IN CATHETER BODY RELATED TO BALLOON INFLATION.
IT WAS REPORTED THAT THERE WAS LEAKAGE OBSERVED FROM THE BALLOON. NO PT COMPLICATIONS WERE REPORTED. ONE CATHETER WAS REPORTED, THREE CATHETERS WERE RETURNED FOR EVAL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SWAN-GANZ CONTINUOUS CARDIAC OUTPUT CATHETER | CONTINUOUS CARDIAC OUTPUT CATHETER | DYG | EDWARDS LIFESCIENCES, PR | 774HF75 | 58345848 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other |