FDA Adverse Event Malfunction Summary report: N

M SERIES

MDR report key: 1030802 · Received April 17, 2008

Report

Report Number
1220908-2008-00762
Event Type
Malfunction
Date Received
April 17, 2008
Report Date
March 31, 2008
Manufacturer
ZOLL MEDICAL CORP
Product Code
MKJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORP HAS NOT REC'D THE DEVICE FOR EVAL AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED A "BRIDGE TEST FAIL" MESSAGE. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORP MSERIES BI-PHASIC NA

Patients

Seq Age Sex Outcome Treatment
1 NA