ANALYTICAL P MODULE
Report
- Report Number
- 1823260-2008-03283
- Event Type
- Malfunction
- Date Received
- April 16, 2008
- Date of Event
- March 25, 2008
- Report Date
- April 16, 2008
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JJE
- PMA / PMN Number
- K953239
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ACCOUNT EXPERIENCING DISCREPANT CREATININE RESULTS FOR 2 PATIENT SAMPLES. PATIENT 1, INITIAL RESULT GAVE 2.96 MG/DL; REPEAT GAVE 1.16 MG/DL. ERRONEOUS RESULT REPORTED ONLY FOR PATIENT 1. PATIENT NOT ADVERSELY AFFECTED. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE CAUSE FOR THE DISCREPANCIES TO BE DUE THE SAMPLE PROBE BLINDING IN SAMPLE PROBE ARM, AND CLEANED SAMPLE PROBE AND PROBE HOLDER. PERFORMANCE TESTS WERE PERFORMED AND WITHIN SPECIFICATION.
ACCOUNT EXPERIENCING DISCREPANT CREATININE RESULTS FOR 2 PATIENT SAMPLES. PATIENT 2: OCCURRED IN 2008, INITIAL RESULT GAVE 0.62 MG/DL; REPEAT GAVE <0.30 MG/DL. ERRONEOUS RESULT REPORTED ONLY FOR PATIENT 1. PATIENT NOT ADVERSELY AFFECTED. THE FIELD SERVICE REPRESENTATIVE DETERMINED THE CAUSE FOR THE DISCREPANCIES TO BE DUE THE SAMPLE PROBE BLINDING IN SAMPLE PROBE ARM, AND CLEANED SAMPLE PROBE AND PROBE HOLDER. PERFORMANCE TESTS WERE PERFORMED AND WITHIN SPECIFICATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANALYTICAL P MODULE | CLINICAL CHEMISTRY ANALYZER - JJE | JJE | ROCHE DIAGNOSTICS | P MODULE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 22 YR | |||
| 2 | 9 DA |