FDA Adverse Event Malfunction Summary report: N

SYRINGE 0.5ML 31GA 8MM 10 BAG 500 WAL

MDR report key: 10305902 · Received July 21, 2020

Report

Report Number
1920898-2020-00909
Event Type
Malfunction
Date Received
July 21, 2020
Date of Event
July 13, 2020
Report Date
July 15, 2020
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
UDI-DI
00681131311762
PMA / PMN Number
K024112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MEDICAL DEVICE EXPIRATION DATE: NA. INVESTIGATION SUMMARY: LEVEL A INVESTIGATION - COMPLAINT EVALUATION / COMPLAINT HISTORY CHECK FOR THE EVENT(S) THAT OCCURRED. SEVERITY: S_1__; OCCURRENCE: A COMPLAINT HISTORY CHECK WAS PERFORMED AND THIS IS THE 1ST RELATED COMPLAINT FOR SHIELD DOES NOT DETACH AS INTENDED AND NEEDLE HUB SEPARATES ON LOT # 9231073. A REVIEW OF RISK MANAGEMENT 150RMN-0001-16 REVISION 13 INDICATES THAT THE POTENTIAL RISK OF THIS SPECIFIC REPORTED INCIDENT (SYRINGE, SHIELD DOES NOT DETACH AS INTENDED, NEEDLE HUB SEPARATES) WAS CAPTURED AND ADDRESSED APPROPRIATELY. INVESTIGATION SUMMARY: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED. IF SAMPLES ARE RECEIVED IN THE FUTURE THE COMPLAINT WILL BE REOPENED FOR FURTHER INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 9231073. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WAS ONE (1) NOTIFICATION [200843441] NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. THERE WAS ONE (1) NOTIFICATION [200842833] NOTED FOR RAISED HUBS. THERE WAS ONE (1) NOTIFICATION [200833554] NOTED FOR OUT OF SPEC SHIELD PULL. INVESTIGATION CONCLUSION: AS NO SAMPLES AND/OR PHOTO(S) WERE RECEIVED THE INVESTIGATION CONCLUDED: - UNCONFIRMED: BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE AS NO SAMPLES OR PHOTOS WERE RETURNED COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. ROOT CAUSE DESCRIPTION: ROOT CAUSE CANNOT BE DETERMINED AT THIS TIME AS THE ISSUE IS UNCONFIRMED AS NO SAMPLES OR PHOTOS WERE RETURNED. RATIONALE: BASED ON THE INVESTIGATION, NO ADDITIONAL INVESTIGATION AND NO CAPA IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT SYRINGE 0.5ML 31GA 8MM 10 BAG 500 WAL HUB SEPARATED DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO. 328509 BATCH NO. 9231073 IT WAS REPORTED THAT NEEDLE SHIELD WAS DIFFICULT TO REMOVE AND NEEDLE HUB SEPARATES AND STAYS INSIDE OF SHIELD. VERBATIM: CONSUMER REPORTED NEEDLE SHIELD DIFFICULT TO REMOVE. ALSO STATED THAT THE NEEDLE HUB SEPARATES AND STAYS INSIDE OF THE SHIELD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
770198 SYRINGE 0.5ML 31GA 8MM 10 BAG 500 WAL PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 328509 9231073 00681131311762

Patients

Seq Age Sex Outcome Treatment
1 Other