FDA Adverse Event Malfunction Summary report: N

M SERIES

MDR report key: 1030490 · Received April 16, 2008

Report

Report Number
1220908-2008-00777
Event Type
Malfunction
Date Received
April 16, 2008
Report Date
April 2, 2008
Manufacturer
ZOLL MEDICAL CORP
Product Code
MKJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORP HAS NOT REC'D THE PRODUCT, AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE DISPLAYED "PACER FAULT 116" AND "DEFIB DISABLED" MESSAGES. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORP MSERIES BI-PHASIC NA

Patients

Seq Age Sex Outcome Treatment
1 NA