FDA Adverse Event
Malfunction
Summary report: N
CONTOUR TS TEST STRIPS (50)
MDR report key: 1030395
·
Received April 15, 2008
Report
- Report Number
- 1826988-2008-00419
- Event Type
- Malfunction
- Date Received
- April 15, 2008
- Date of Event
- March 27, 2008
- Report Date
- March 27, 2008
- Manufacturer
- BAYER HEALTHCARE, LLC
- Product Code
- NBW
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER STATED THAT SHE TESTED HER BLOOD GLUCOSE USING HER CONTOUR METER AND ANOTHER METER. THE CONTOUR READ 501 MG/DL, WHILE THE OTHER METER READ 91 MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "D" ZONE OF THE PARKES ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. WHILE TROUBLESHOOTING, THE CUSTOMER PERFORMED A CONTROL TEST AND RECEIVED A RESULT OF 218 MG/DL. THE NORMAL CONTROL RANGE WAS 99-136 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS ARE TO BE RETURNED FOR EVALUATION. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR TS TEST STRIPS (50) | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE, LLC | 1820 | 7LD3E07 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |