FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS (50)

MDR report key: 1030370 · Received April 15, 2008

Report

Report Number
1826988-2008-00431
Event Type
Malfunction
Date Received
April 15, 2008
Date of Event
March 30, 2008
Report Date
March 30, 2008
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER CALLED FOR HELP WITH HIS METER. WHILE TROUBLESHOOTING, HE PERFORMED CONTROL TESTS AND RECEIVED A RESULT OF 194 MG/DL. THE NORMAL CONTROL RANGE WAS 101-139 MG/DL. NO ADVERSE EVENTS WERE ALLEGED. THE TEST STRIPS WERE RETURNED FOR EVAL. REPLACEMENT TEST STRIPS WERE SENT TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS (50) BLOOD GLUCOSE TEST STRIPS NBW BAYER HEALTHCARE, LLC NOT PROVIDED 7EC3B02

Patients

Seq Age Sex Outcome Treatment
1 UNK