FDA Adverse Event Malfunction Summary report: N

PULSE GEN MODEL 102

MDR report key: 1030362 · Received April 16, 2008

Report

Report Number
1644487-2008-00934
Event Type
Malfunction
Date Received
April 16, 2008
Date of Event
March 17, 2008
Report Date
March 17, 2008
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

REPORTER INDICATED THAT COMMUNICATION COULD NOT BE ESTABLISHED WITH A VNS PT'S GENERATOR. COMMUNICATION WAS ATTEMPTED WITH MULTIPLE PROGRAMMING SYSTEMS. A BATTERY LIFE ESTIMATION WAS PERFORMED, AND IT RESULTED IN 4.65 YEARS REMAINING UNTIL ERI=YES, ALTHOUGH THE VALUES USED MAY NOT BE REPRESENTATIVE OF THE PT'S EXACT PROGRAMMED PARAMETERS. THE PT UNDERWENT REVISION SURGERY. THE SURGEON NOTED THAT DURING THE REVISION THERE WAS SIGNIFICANT CORROSION AT THE GENERATOR HEADER/LEAD CONNECTOR PIN INTERFACE. THIS LED THE SURGEON TO REPLACE THE GENERATOR AND LEAD. THE EXPLANTED DEVICES HAVE BEEN RECEIVED AND ARE AWAITING ANALYSIS.