NEUROSIGN
Report
- Report Number
- 8021774-2020-00004
- Event Type
- Malfunction
- Date Received
- July 21, 2020
- Report Date
- June 22, 2020
- Manufacturer
- THE MAGSTIM COMPANY LTD.
- Product Code
- ETN
- PMA / PMN Number
- K923056
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION RESULTS: THE POTENTIOMETERS (22K LOG 475048), VR1 ON THE 2527-01 FRONT PANEL PCB AND VR1 ON THE 2528-01 PIGGY BACK PCBARE WORN AND REQUIRE REPLACEMENT. THE WORN POTENTIONMETER IS THE ROOT CAUSE. TREND ANALYSIS: A REVIEW WAS PERFORMED OF SIMILAR INCIDENTS, FROM 01 JANUARY 2017 TO 22 JUNE 2020 (AWARENESS DATE). ONLY ONE SIMILAR EVENT WAS NOTED, WITH OUR REFERENCE CC 954. HOWEVER, UPON RE-ASSESSMENT OF THE RISK ASSOCIATED WITH THIS EVENT (CC 954), IT WAS CONCLUDED THAT THE DEVICE HAD EXCEEDED ITS LIFETIME. OUT OF THE LIFETIME OF A DEVICE, IT CAN BE REASONABLY EXPECTED TO START TO SHOW WEAR AND TEAR. A TREND IN A FAULTY/WORN POTENTIOMETER IN THIS DEVICE WITHIN ITS LIFETIME, IS NOT CURRENTLY SHOWING. WE WILL KEEP MONITORING THIS IN LINE WITH OUR COMPLAINT HANDLING PROCEDURE AND TAKE CORRECTIVE AND PREVENTIVE ACTIONS WHERE REQUIRED. NOTE: WE WERE INFORMED THAT THE ISSUE DID NOT OCCUR IN DENMARK BUT IN SWEDEN. THE DISTRIBUTOR IS BASED IN DENMARK, AND THE CUSTOMER IN SWEDEN.
THE DEVICE WAS RETURNED FOR INVESTIGATION. THERE IS EVIDENCE TO SUGGEST THAT THE POTENTIOMETER WAS THE CAUSE OF THE ISSUE. FURTHER INVESTIGATION IS ONGOING.
IT WAS REPORTED THAT THE STIMULATOR PRESENTED A VERY WEAK SOUND, EVEN WITH THE HIGHEST CURRENT. NO ACTUAL INCIDENTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 765242 | NEUROSIGN | NEUROSIGN 100 | ETN | THE MAGSTIM COMPANY LTD. | 100 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | 1020-00 MUTE SENSOR - SN (B)(6).| 1719-00 N100 PREAMPLIFIER POD - SN (B)(6).| 9967-00 STIMULATOR POD - SN (B)(6). |