FULL FACE MASK
Report
- Report Number
- 9611451-2008-00186
- Event Type
- Malfunction
- Date Received
- April 15, 2008
- Date of Event
- March 6, 2008
- Report Date
- March 8, 2008
- Manufacturer
- FISHER & PAYKEL HEALTHCARE, LTD.
- Product Code
- MNT
- PMA / PMN Number
- K060044
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE RETURNED DEVICE WAS VISUALLY INSPECTED. RESULTS: THE SEAL HAD SEPARATED FROM THE MASK. THIS IS A KNOWN FAILURE MODE. A LOT CHECK HAS REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR THIS LOT NUMBER. THE SEAL IS HELD ON THE MASK BASE BY FRICTION BETWEEN THE SILICONE SEAL AND THE RETAINING SLOT. SEVERAL UNDERCUT BUMP FEATURES PROVIDE SOME MECHANICAL LOCKING. HOWEVER THE SEAL CAN BE PULLED OUT UNDER A FORCE OF APPROXIMATELY 10-15N. THIS 'MECHANICAL LOCKING' HAS BEEN DEEMED TO BE INSUFFICIENT AS COMPLAINTS HAVE BEEN REPORTED DUE TO THE SEAL (CUSHION) COMING OFF DURING USE OR DURING TRANSIT. THE MANUFACTURING PROCESS HAS BEEN UPDATED AND A NEW GLUING PROCESS HAS BEEN ADDED TO PROVIDE BETTER SEAL RETENTION. THE NEW PROCESS HAS COME INTO EFFECT FROM 5 MARCH 2008. OUR MONITORING AND TRENDING FOR THIS TYPE OF EVENT SHOWS A RATE OF OCCURRENCE OF 0.0531% WORLDWIDE FOR THE PAST YEAR.
A HEALTHCARE FACILITY REPORTED TO OUR DISTRIBUTOR THAT THE SEAL OF A FULL FACE MASK HAD SEPARATED FROM THE MASK BASE ON THE REGION OF THE MASK UNDER THE CHIN. THIS WAS REPORTED TO BE AN OUT OF PACKAGE FAILURE. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FULL FACE MASK | MNT | FISHER & PAYKEL HEALTHCARE, LTD. | RT040L | 071120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |