FDA Adverse Event Malfunction Summary report: N

FULL FACE MASK

MDR report key: 1030339 · Received April 15, 2008

Report

Report Number
9611451-2008-00171
Event Type
Malfunction
Date Received
April 15, 2008
Date of Event
March 6, 2008
Report Date
March 6, 2008
Manufacturer
FISHER & PAYKEL HEALTHCARE, LTD.
Product Code
MNT
PMA / PMN Number
K060044
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

METHOD: AN INVESTIGATION WAS CONDUCTED ON THE RETURNED UNIT. RESULTS: THE DEVICE WAS BROKEN AT THE NOSE BRIDGE. CONCLUSION: STRENGTH TESTING OF THE FOREHEAD PIECE HAS SHOWN THAT IT TAKES AN AVERAGE FORCE OF 63.7 N (6.4KG) FOR STRESS MARKS TO APPEAR AND AN AVERAGE FORCE OF 82.6 N (8.4KG) FOR FAILURE TO OCCUR. IT TAKES AN EXCESSIVE FORCE ON THE FOREHEAD REST FOR FAILURE. THE CAUSE OF THE FAILURE WAS THE APPLICATION OF EXCESSIVE FORCE, BUT IT IS NOT POSSIBLE TO DETERMINE THE SOURCE AND AMOUNT OF THE FORCE THAT CAUSED THIS FAILURE. OUR MONITORING AND TRENDING FOR THIS TYPE OF EVENT SHOWS A RATE OF OCCURRENCE OF 0.0004% WORLDWIDE FOR THE PAST YEAR.

Description of Event or Problem · 1

A HEALTHCARE FACILITY REPORTED TO OUR DISTRIBUTOR THAT THERE WAS A BREAK IN THE PLASTIC AT THE NASAL BRIDGE OF AN RT040 FULL FACE MASK. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FULL FACE MASK MNT FISHER & PAYKEL HEALTHCARE, LTD. RT040L 071115

Patients

Seq Age Sex Outcome Treatment
1