FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 1030311
·
Received April 15, 2008
Report
- Report Number
- 1811755-2008-00021
- Event Type
- Malfunction
- Date Received
- April 15, 2008
- Date of Event
- March 14, 2008
- Report Date
- March 14, 2008
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- PHYSICIAN ASSISTANT
Narratives
Additional Manufacturer Narrative · 1
DURING THE INVESTIGATION, SIGNIFICANT CORROSION WAS LOCATED INSIDE THE HANDPIECE. THE BEARINGS AND DRIVESHAFT BOTH HAD TO BE REPLACED AS A RESULT OF THE CORROSION. FURTHER INVESTIGATION DETERMINED THAT THE BUR GUARD HAD A SEIZED BEARING WITH ADD'L SIGNS OF CORROSION. BASED OFF OF THE INVESTIGATION FINDINGS, IMPROPER STERILIZATION PROCEDURES AT THE CUSTOMER ACCOUNT WAS LIKELY TO BE THE CONTRIBUTING FACTOR FOR THE HANDPIECE AND THE BUR GUARD TO FUNCTION IMPROPERLY.
Description of Event or Problem · 1
PER THE SALES REP, DURING A WISDOM TOOTH REMOVAL, THE PT RECEIVED A MINOR BURN TO THE LOWER RIGHT PORTION OF THE LIP. NO MEDICAL INTERVENTION WAS REQUIRED TO TREAT THE BURN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | CORE SYSTEM | DZJ | STRYKER INSTRUMENTS KALAMAZOO | 0604009793 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |