DIMENSION CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 1226181-2008-00039
- Event Type
- Malfunction
- Date Received
- April 15, 2008
- Date of Event
- March 18, 2008
- Report Date
- March 18, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC.
- Product Code
- MMI
- PMA / PMN Number
- K9440093
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
NO FURTHER EVAL OF THE DEVICE IS REQUIRED. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULTS WAS A LOOSE HM WASH PROBE. THE MALFUNCTION WAS RESOLVED AFTER THE CUSTOMER CORRECTED THE PROBLEM WITH GUIDANCE FROM A SIEMENS HEALTHCARE DIAGNOSTICS, INC TECHNICAL SOLUTION CENTER REPRESENTATIVE. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.
FALSELY ELELVATED TROPONIN I RESULTS WERE OBTAINED ON 3 PTS' SAMPLES. THE RESULTS WERE REPORTED TO THE PHYSICIAN. THE SAMPLES WERE RETESTED AND NEGATIVE RESULTS WERE OBTAINED. THE PTS WERE ADMITTED TO THE HOSPITAL. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULTS. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELELVATED TROPONIN RESULTS WAS A LOOSE HM WASH PROBE.
NO FURTHER EVAL OF THE DEVICE IS REQUIRED. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULTS WAS A LOOSE HM WASH PROBE. THE MALFUNCTION WAS RESOLVED AFTER THE CUSTOMER CORRECTED THE PROBLEM WITH GUIDANCE FROM A SIEMENS HEALTHCARE DIAGNOSTICS, INC TECHNICAL SOLUTION CENTER REPRESENTATIVE. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.
FALSELY ELELVATED TROPONIN I RESULTS WERE OBTAINED ON 3 PTS' SAMPLES. THE RESULTS WERE REPORTED TO THE PHYSICIAN. THE SAMPLES WERE RETESTED AND NEGATIVE RESULTS WERE OBTAINED. THE PTS WERE ADMITTED TO THE HOSPITAL. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULTS. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELELVATED TROPONIN RESULTS WAS A LOOSE HM WASH PROBE.
NO FURTHER EVAL OF THE DEVICE IS REQUIRED. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULTS WAS A LOOSE HM WASH PROBE. THE MALFUNCTION WAS RESOLVED AFTER THE CUSTOMER CORRECTED THE PROBLEM WITH GUIDANCE FROM A SIEMENS HEALTHCARE DIAGNOSTICS, INC TECHNICAL SOLUTION CENTER REPRESENTATIVE. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.
FALSELY ELELVATED TROPONIN I RESULTS WERE OBTAINED ON 3 PTS' SAMPLES. THE RESULTS WERE REPORTED TO THE PHYSICIAN. THE SAMPLES WERE RETESTED AND NEGATIVE RESULTS WERE OBTAINED. THE PTS WERE ADMITTED TO THE HOSPITAL. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULTS. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELELVATED TROPONIN RESULTS WAS A LOOSE HM WASH PROBE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION CLINICAL CHEMISTRY SYSTEM | CLINICAL CHEMISTRY SYSTEM | MMI | SIEMENS HEALTHCARE DIAGNOSTICS, INC. | RXLMAXRH | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ||||
| 2 | ||||
| 3 |