STRATUS CS STAT FLUOROMETRIC ANALYZER
Report
- Report Number
- 2517506-2008-00035
- Event Type
- Malfunction
- Date Received
- April 15, 2008
- Date of Event
- March 15, 2008
- Report Date
- March 17, 2008
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS, INC.
- Product Code
- MMI
- PMA / PMN Number
- K051650
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS HETEROPHILIC ANTIBODY. THE IFU FOR STRATUS CS CCTIN TESTPAK CONTAINS THE FOLLOWING INFORMATION: "PATIENT SAMPLES MAY CONTAIN HETEROPHILIC ANTIBODIES THAT COULD REACT IN IMMUNOASSAYS TO GIVE FALSELY ELEVATED OR DEPRESSED RESULTS. THIS ASSAY HAS BEEN DESIGNED TO MINIMIZE INTERFERENCE FROM HETEROPHILIC ANTIBODIES. NEVERTHELESS, COMPLETE ELIMINATION OF THIS INTERFERENCE CANNOT BE GUARANTEED. A TEST RESULT THAT IS INCONSISTENT WITH THE CLINICAL PICTURE AND PATIENT HISTORY SHOULD BE INTERPRETED WITH CAUTION." THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS.
A FALSELY ELEVATED TROPONIN I RESULT OF 1.74 NG/ML WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS REPORTED TO THE PHYSICIAN WHO QUESTIONED THE RESULT. THE ORIGINAL SAMPLE WAS RETESTED ON ALTERNATE METHODOLOGY AND A RESULT OF 0.00 NG/ML WAS OBTAINED. PATIENT TREATMENT WAS NOT PRESCRIBED OR ALTERED AN THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT. ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS HETEROPHILIC ANTIBODY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STRATUS CS STAT FLUOROMETRIC ANALYZER | ACUTE CARE TROPONIN I TEST PAK | MMI | SIEMENS HEALTHCARE DIAGNOSTICS, INC. | NA | 238021002 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |