FDA Adverse Event Injury Summary report: N

NXSTAGE SYSTEM ONE

MDR report key: 1030162 · Received April 16, 2008

Report

Report Number
3003464075-2008-00169
Event Type
Injury
Date Received
April 16, 2008
Date of Event
March 18, 2008
Report Date
March 18, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED CARTRIDGE IS UNDER INVESTIGATION AT THE TIME OF THIS REPORT. THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT PERFORMING RINSEBACK AS INSTRUCTED IN THE USER'S GUIDE. THE USER'S GUIDE STATES TO CHECK THE WASTE FLUID FOR THE PRESENCE OF BLOOD, IF PINK TINGED TERMINATE TREATMENT AND RINSEBACK BLOOD. A BLOOD LOSS SHOULD NOT OCCUR IF DEVICE LABELING IS FOLLOWED. FACILITY STAFF HAS BEEN NOTIFIED. THERE HAVE BEEN NO SIMILAR PROBLEMS REPORTED SUBSEQUENT TO THIS EVENT. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADDITIONAL INFORMATION WILL BE PROVIDED.

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AS AN ADVERSE EVENT SOLELY TO COMPLY WITH THE FDA'S BLOOD LOSS POLICY. A BLOOD LEAK DETECTOR ALARM OCCURRED DURING A ROUTINE HEMODIALYSIS TREATMENT. THE OPERATOR NOTED THE EFFLUENT WAS PINK. TREATMENT WAS DISCONTINUED WITHOUT PERFORMING RINSEBACK, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 8017701

Patients

Seq Age Sex Outcome Treatment
1 54 YR Other