NOVASURE IMPEDANCE CONTROLLED EA SYSTEM
Report
- Report Number
- 1222780-2008-00040
- Event Type
- Injury
- Date Received
- April 16, 2008
- Date of Event
- March 18, 2008
- Report Date
- March 18, 2008
- Manufacturer
- CYTYC SURGICAL PRODUCTS
- Product Code
- MNB
- PMA / PMN Number
- P010013
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE HISTORY RECORD (DHR) REVIEW WAS CONDUCTED FOR THE REPORTED LOT NUMBER. THE LOT WAS RELEASED MEETING ALL QA SPECIFICATIONS. CURRENTLY UNABLE TO ESTABLISH A RELATIONSHIP OR IMPACT TO THE REPORTED OBSERVATION. ACCORDING TO THE IFU UNDER THE WARNINGS SECTION: USE CAUTION NOT TO PERFORATE THE UTERINE WALL WHEN SOUNDING, DILATING, OR INSERTING THE DISPOSABLE DEVICE. IF THE DISPOSABLE DEVICE IS DIFFICULT TO INSERT INTO THE CERVICAL CANAL, USE CLINICAL JUDGMENT TO DETERMINE WHETHER OR NOT FURTHER DILATION IS REQUIRED. THE NOVASURE SYSTEM PERFORMS A CAVITY INTEGRITY ASSESSMENT (CIA) TEST TO EVALUATE THE INTEGRITY OF THE UTERINE CAVITY, AND SOUNDS AN ALARM WARNING OF A POSSIBLE PERFORATION PRIOR TO TREATMENT. (STEP 2.36) ALTHOUGH DESIGNED TO DETECT A PERFORATION OF THE UTERINE WALL, IT (CIA ALARM) IS AN INDICATOR ONLY AND IT MIGHT NOT DETECT ALL PERFORATIONS UNDER ALL POSSIBLE CIRCUMSTANCES. CLINICAL JUDGEMENT MUST ALWAYS BE USED.
USER FACILITY REPORTED 2 UNSUCCESSFUL CAVITY ASSESSMENT INTEGRITY (CIA) TESTS WITH A DISPOSABLE NOVASURE DEVICE. ON HYSTEROSCOPY, A UTERINE PERFORATION WAS SEEN. NO TREATMENT WAS NEEDED AND THE PT WAS DISCHARGED HOME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NOVASURE IMPEDANCE CONTROLLED EA SYSTEM | RADIO FREQUENCY ENDOMETRIAL ABLATION | MNB | CYTYC SURGICAL PRODUCTS | NS2000 | 07LO3HB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | Other |