SPINNING SPIROS CLOSED MALE LUER, RED CAP
Report
- Report Number
- 9617594-2020-00255
- Event Type
- Malfunction
- Date Received
- July 21, 2020
- Date of Event
- June 1, 2020
- Report Date
- June 25, 2020
- Manufacturer
- ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V.
- Product Code
- FPA
- UDI-DI
- 00840619023386
- PMA / PMN Number
- K082806
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHARMACIST
Narratives
NO PRODUCT SAMPLES WERE RETURNED FOR INVESTIGATION, HOWEVER, A PHOTOGRAPH WAS RETURNED SHOWING A 20130-01 SPINNING SPIROS CONNECTED TO A LUER LOCK SYRINGE AND VISIBLE LEAKAGE INTERNAL TO THE 20130-01 SPINNING SPIROS. THE DEVICE HISTORY REVIEW (DHR) WAS REVIEWED AND THERE WERE NO RELEVANT NON-CONFORMANCES FOUND THAT WOULD HAVE CONTRIBUTED TO THE REPORTED COMPLAINT. THOUGH LEAKAGE CAN BE CONFIRMED, A PROBABLE CAUSE CANNOT BE IDENTIFIED WITHOUT THE RETURN OF THE AFFECTED SAMPLE.
THE EVENT OCCURRED BETWEEN (B)(6) 2020 AND INVOLVED A SPINNING SPIROS CLOSED MALE LUER, RED CAP WHERE THE SPIROS SANK AND DID NOT REMAIN STILL OR TILTED DURING AN INFUSION OF DOXORUBICIN USING A 12 CC SYRINGE RESULTING IN A LEAK. THERE WAS NO APPARENT PHYSICAL DEFECT OBSERVED ON THE SPIROS. THERE WAS PATIENT INVOLVEMENT AND NO PATIENT HARM. THIS IS TWO OF TWO INCIDENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 769496 | SPINNING SPIROS CLOSED MALE LUER, RED CAP | SET, ADMINISTRATION, INTRAVASCULAR | FPA | ICU MEDICAL DE MEXICO, S. DE R.L. DE C.V. | 4749754 | 00840619023386 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |