FDA Adverse Event Injury Summary report: N

AUTOMATED PD SET W/CASSETTE4 PRONG

MDR report key: 1030074 · Received April 15, 2008

Report

Report Number
1423500-2008-00235
Event Type
Injury
Date Received
April 15, 2008
Date of Event
March 31, 2008
Report Date
April 3, 2008
Manufacturer
BAXTER HEALTHCARE
Product Code
FKX
PMA / PMN Number
K923065
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL SAMPLE WAS DISCARDED. THE LOT NUMBER IS UNK. BAXTER IS MONITORING ISSUES LIKE THIS. SHOULD ADD'L INFO BECOME AVAILABLE, A FOLLOW UP REPORT WILL BE FILED.

Description of Event or Problem · 1

INITIALLY, THE CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER ON 3-17-08 REGARDING A SYSTEM ERROR 2240 ALARM THAT APPEARED ON THE HOME CHOICE (HC) DISPLAY DURING AN INITIAL DRAIN. A 2240 ALARM IS AN AIR IN THE ALARM. THE PT LINE WAS NOT CONNECTED TO THE TRANSFER SET BEFORE PRESSING, "GO" TO BEGIN THE INITIAL DRAIN/THERAPY. REPORTEDLY, THERE WAS A LEAK NOTED IN THE TUBING SET. NO PT INJURY OR INTERVENTION WAS REPORTED AT THE TIME OF THIS REPORT. DURING A FOLLOW UP CALL ON 4-3-08 WITH THE HOME PT'S NURSE REGARDING THE SYSTEM ERROR 2240 ALARM, THE NURSE INDICATED THAT THE HOME PT WAS HOSPITALIZED DUE TO PERITONITIS. THE HOSPITAL ADVISED THE HOME PT HAD BEEN HOSPITALIZED IN 2008 WITH ABDOMINAL PAIN AND CLOUDY FLUID. THE HOME PT WAS DIAGNOSED WITH PERITONITIS THE NEXT DAY. THE PT'S CULTURE WAS POSITIVE FOR STAPHYLOCOCCUS AND THE INITIAL WHITE BLOOD COUNT (WBC) WAS 10. CEFTRIAXONE 1GM DAILY WAS ORDERED. THE HOME PT WAS DISCHARGED FROM THE HOSPITAL THE NEXT MONTH. THE PT OUTCOME IS GOOD AND HE HAS RESUMED PD THERAPY. THE NURSE INDICATED THE PERITONITIS WAS REPORTEDLY CAUSED BY POOR ASEPTIC TECHNIQUE. HOWEVER, THE PT IS COMPLIANT WITH THERAPY. REPORTEDLY, THE PT HAS A PET THAT CHEWED ON THE TUBING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOMATED PD SET W/CASSETTE4 PRONG 78FKX FKX BAXTER HEALTHCARE UNK

Patients

Seq Age Sex Outcome Treatment
1 38 YR Required Intervention