FDA Adverse Event Injury Summary report: N

NXSTAGE SYSTEM ONE

MDR report key: 1030064 · Received April 15, 2008

Report

Report Number
3003464075-2008-00168
Event Type
Injury
Date Received
April 15, 2008
Date of Event
March 17, 2008
Report Date
March 17, 2008
Manufacturer
NXSTAGE MEDICAL, INC.
Product Code
KDI
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED BLOOD LOSS IS ATTRIBUTED TO THE OPERATOR NOT PERFORMING RINSEBACK AS AIR WAS VISIBLE IN CIRCUIT INDICATING THE CYCLER ALARMED APPROPRIATELY. THE SOURCE OF THE AIR WAS NOT APPARENT TO THE OPERATOR AND THE DISPOSABLE CARTRIDGE WAS NOT AVAILABLE FOR EVAL. THE EXACT CAUSE OF THE AIR COULD NOT BE DETERMINED. THE USER'S GUIDE PROVIDES ADEQUATE INSTRUCTIONS FOR TROUBLESHOOTING AIR ALARMS. A DIRECT CORRELATION BETWEEN NXSTAGE SYSTEM ONE AND THE REPORTED BLOOD LOSS CANNOT BE ESTABLISHED. FACILITY STAFF HAS BEEN NOTIFIED AND WILL REVIEW THE EVENT WITH THE OPERATOR. NXSTAGE MEDICAL CONSIDERS THIS REPORT CLOSED. NO ADD'L INFO WILL BE PROVIDED.

Description of Event or Problem · 1

AN ARTERIAL AIR ALARM OCCURRED DURING A ROUTINE HEMODIALYSIS TREATMENT, WHICH COULD NOT BE RESOLVED BY THE OPERATOR. RINSEBACK WAS NOT PERFORMED, RESULTING IN AN ESTIMATED BLOOD LOSS OF 190CC. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NXSTAGE SYSTEM ONE HIGH PERMEABILITY HEMODIALYSIS SYSTEM KDI NXSTAGE MEDICAL, INC. CAR-170 UNK

Patients

Seq Age Sex Outcome Treatment
1 41 YR Other