FDA Adverse Event Malfunction Summary report: N

640G INSULIN PUMP MMT-1711K

MDR report key: 10300274 · Received July 21, 2020

Report

Report Number
2032227-2020-149565
Event Type
Malfunction
Date Received
July 21, 2020
Date of Event
May 12, 2020
Report Date
May 6, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
00643169739468
Removal / Correction Number
Z-0958-2020
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS PART OF A RETROSPECTIVE REVIEW AND REMEDIATION EFFORTS IN RESPONSE TO A WARNING LETTER. UPDATED H9: Z-0958-2020. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 1

PUMP PASSED SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, SELF TEST AND DISPLACEMENT TEST. HOWEVER, PUMP TRACE DOWNLOAD ANALYSIS CONFIRMED THE PUMP ALARMED POWER ERROR DETECTED, LOW BATTERY ALERT, POWER LOSS ALARM AND REPLACE BATTERY ALERT/REPLACE BATTERY NOW ALARM. THE POWER MANAGEMENT TOOL CONFIRMED POWER ERROR DETECTED, LOW BATTERY ALERT, POWER LOSS ALARM AND REPLACE BATTERY ALERT/REPLACE BATTERY NOW ALARM WAS TRIGGERED WHEN THE BACKUP BATTERY LOADED VOLTAGE (LOADED VLITH) WAS LESS THAN 3.5 V FOR 4 CONSECUTIVE HOURS DUE TO CONNECTOR RESISTANCE ELECTRICAL BOARD. AFTER DISCONNECTING AND RECONNECTING THE INTERNAL BATTERY CONNECTOR ON ELECTRICAL BOARD, THE PUMP WAS MONITORED AND FUNCTIONED PROPERLY. (B)(4). THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE 640 G INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED REPLACED BATTERY NOW ALARM. CUSTOMER STATED THE BATTERY CAP WAS NOT LOOSE, CRACKED OR DAMAGED. THIS WAS THE SECOND OCCURRENCE OF REPLACE BATTERY ALERT OR REPLACE BATTERY NOW ALARM IN LESS THAN 8 HOURS FROM A LOW BATTERY WARNING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
768029 640G INSULIN PUMP MMT-1711K PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1711K HG1MMJ8 00643169739468

Patients

Seq Age Sex Outcome Treatment
1 Unknown