FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II CLOSED IV CATHETER SYSTEM

MDR report key: 10300043 · Received July 20, 2020

Report

Report Number
3006948883-2020-00300
Event Type
Malfunction
Date Received
July 20, 2020
Date of Event
June 21, 2020
Report Date
August 12, 2020
Manufacturer
BD (SUZHOU)
Product Code
FOZ
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION SUMMARY : A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 8171142. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, VISUAL EXAMINATION WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT. THE RETENTION SAMPLES WERE FOUND TO BE FREE OF ANY ABNORMALITIES OR DEFECTS THAT COULD CONTRIBUTE TO A BROKEN CANNULA. UNFORTUNATELY WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD INTIMA-II¿ CLOSED IV CATHETER SYSTEM EXPERIENCED A BROKEN/DETACHED NEEDLE. PRODUCT DEFECT WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ON (B)(6) 2020, A BROKEN NEEDLE WAS FOUND DURING INTRAVENOUS INFUSION WITH INDWELLING NEEDLE.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD INTIMA-II CLOSED IV CATHETER SYSTEM EXPERIENCED A BROKEN/DETACHED NEEDLE. PRODUCT DEFECT WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: ON (B)(6) 2020, A BROKEN NEEDLE WAS FOUND DURING INTRAVENOUS INFUSION WITH INDWELLING NEEDLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
764198 BD INTIMA-II CLOSED IV CATHETER SYSTEM INTERVASCULAR CATHETER FOZ BD (SUZHOU) 8171142

Patients

Seq Age Sex Outcome Treatment
1 Other