SKELETAL DYNAMICS
Report
- Report Number
- 3006742481-2019-00002
- Event Type
- Malfunction
- Date Received
- July 20, 2020
- Date of Event
- June 25, 2019
- Report Date
- October 9, 2019
- Manufacturer
- SKELETAL DYNAMICS
- Product Code
- HWC
- UDI-DI
- 00841506101286
- PMA / PMN Number
- K182492
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
UPON RECEIVING THE FDA REPORT (MW5089046), A CALL WAS MADE TO THE REPRESENTATIVE TO GET MORE INFORMATION FORM THE COMPLAINT. THE INFORMATION PROVIDED WAS THAT THE SURGEON WAS HAVING ISSUES WITH THE VARIABLE ANGLE DRIVER TIP DURING FINAL TORQUEING. WE WERE UNABLE TO GET MORE INFORMATION REGARDING THIS EVENT VIA THE REPRESENTATIVE. THE INSTRUMENT WAS NOT RETURNED TO BE INVESTIGATED. THE INFORMATION PROVIDED IS INSUFFICIENT TO CONCLUDE A CAUSATION OF WHAT ACTUALLY HAPPENED. TESTING WAS PERFORMED ON A SIMILAR DRIVER MODEL IN STOCK AND WAS SEEN TO BE WITHIN SPEC OF THE ALLOTTED TORQUE LIMITS.
USER SUBMITTED A REPORT TO THE FDA STATING: WHILE PLATING A VARIABLE ANGLE SCREW, THE TIP OF THE DRIVER COLD WELDED IN THE HEAD OF THE SCREW AND BROKE OFF. DECISION WAS MADE TO LEAVE THE TIP OF THE DRIVER IN THE SCREW HEAD, AS SCREW HAD GOOD PLACEMENT AND TIP WAS COLD WELDED WITHOUT POSSIBILITY OF MIGRATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 763958 | SKELETAL DYNAMICS | DRIVER, AO CONNECTION, POLYAXIAL LOCKING SCREW | HWC | SKELETAL DYNAMICS | Driver, AO Connection, Polyaxial Locking Screw | 00841506101286 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |