FDA Adverse Event Malfunction Summary report: N

SKELETAL DYNAMICS

MDR report key: 10299817 · Received July 20, 2020

Report

Report Number
3006742481-2019-00002
Event Type
Malfunction
Date Received
July 20, 2020
Date of Event
June 25, 2019
Report Date
October 9, 2019
Manufacturer
SKELETAL DYNAMICS
Product Code
HWC
UDI-DI
00841506101286
PMA / PMN Number
K182492
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIVING THE FDA REPORT (MW5089046), A CALL WAS MADE TO THE REPRESENTATIVE TO GET MORE INFORMATION FORM THE COMPLAINT. THE INFORMATION PROVIDED WAS THAT THE SURGEON WAS HAVING ISSUES WITH THE VARIABLE ANGLE DRIVER TIP DURING FINAL TORQUEING. WE WERE UNABLE TO GET MORE INFORMATION REGARDING THIS EVENT VIA THE REPRESENTATIVE. THE INSTRUMENT WAS NOT RETURNED TO BE INVESTIGATED. THE INFORMATION PROVIDED IS INSUFFICIENT TO CONCLUDE A CAUSATION OF WHAT ACTUALLY HAPPENED. TESTING WAS PERFORMED ON A SIMILAR DRIVER MODEL IN STOCK AND WAS SEEN TO BE WITHIN SPEC OF THE ALLOTTED TORQUE LIMITS.

Description of Event or Problem · 1

USER SUBMITTED A REPORT TO THE FDA STATING: WHILE PLATING A VARIABLE ANGLE SCREW, THE TIP OF THE DRIVER COLD WELDED IN THE HEAD OF THE SCREW AND BROKE OFF. DECISION WAS MADE TO LEAVE THE TIP OF THE DRIVER IN THE SCREW HEAD, AS SCREW HAD GOOD PLACEMENT AND TIP WAS COLD WELDED WITHOUT POSSIBILITY OF MIGRATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
763958 SKELETAL DYNAMICS DRIVER, AO CONNECTION, POLYAXIAL LOCKING SCREW HWC SKELETAL DYNAMICS Driver, AO Connection, Polyaxial Locking Screw 00841506101286

Patients

Seq Age Sex Outcome Treatment
1 Other