FDA Adverse Event Malfunction Summary report: N

BD Q-SYTE LUER ACCESS SPLIT-SEPTUM STAND-ALONE DEVICE

MDR report key: 10299756 · Received July 20, 2020

Report

Report Number
9610847-2020-00212
Event Type
Malfunction
Date Received
July 20, 2020
Date of Event
June 28, 2020
Report Date
August 14, 2020
Manufacturer
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
Product Code
FPA
UDI-DI
30382903851004
PMA / PMN Number
K013621
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: OUR QUALITY ENGINEER REVIEWED THE PHOTOGRAPHS SUBMITTED FOR EVALUATION. UPON INSPECTION OF THE CUSTOMER PHOTOS LEAKAGE CAN BE OBSERVED COMING THROUGH ONE OF THE VENT HOLES ON THE Q-SYTE BODY. THIS LEAKAGE INDICATES DAMAGE TO THE SEPTUM INSIDE OF THE BODY. ALTHOUGH THE REPORTED ISSUE WAS CONFIRMED, IT COULD NOT BE DETERMINED IF THE DAMAGE RESULTED FROM THE MANUFACTURING OF THE DEVICE OR FROM THE USER ENVIRONMENT. A DEVICE HISTORY RECORD REVIEW SHOWED NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF THIS BATCH. SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD Q-SYTE¿ LUER ACCESS SPLIT-SEPTUM STAND-ALONE DEVICE LEAKED AT THE HOLE OF THE SEPTUM DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM CHINESE TO ENGLISH: "DURING THE USE OF Q-SYTE, LEAKAGE WAS FOUND AT THE HOLE OF THE SEPTUM."

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD Q-SYTE¿ LUER ACCESS SPLIT-SEPTUM STAND-ALONE DEVICE LEAKED AT THE HOLE OF THE SEPTUM DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, TRANSLATED FROM (B)(6) TO ENGLISH: "DURING THE USE OF Q-SYTE, LEAKAGE WAS FOUND AT THE HOLE OF THE SEPTUM."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
764169 BD Q-SYTE LUER ACCESS SPLIT-SEPTUM STAND-ALONE DEVICE INTRAVASCULAR ADMINISTRATION SET FPA BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. 385100 9122580 30382903851004

Patients

Seq Age Sex Outcome Treatment
1 Other