FDA Adverse Event Malfunction Summary report: N

CRANIOTOMY PACK (CNKDC)692

MDR report key: 10299197 · Received July 20, 2020

Report

Report Number
1423537-2020-00486
Event Type
Malfunction
Date Received
July 20, 2020
Date of Event
June 24, 2020
Report Date
July 20, 2020
Manufacturer
MEX03 MEXICO-JUAREZ PRESOURCE
Product Code
OJG
UDI-DI
10888439466560
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

BASED ON THE SUPPLIER¿S INVESTIGATION, DEVICE HISTORY RECORD COULD NOT BE REVIEWED BECAUSE THE LOT NUMBER WAS NOT PROVIDED. HOWEVER, THE SUPPLIER REVIEWED THE RECORDS IN THE LAST TWO YEARS AND NO ABNORMAL SITUATION WAS FOUND. NO SAMPLE WAS AVAILABLE AT THE TIME OF THE INVESTIGATION. ACCORDING TO SUPPLIER, OR TOWEL IS MADE OF COTTON, SO COTTON FIBER IS BORN. THE SUPPLIER IS CONTINUOUSLY WORKING WITH CARDINAL HEALTH TO BETTER CONTROL THE LINTING AND HAVE IMPLEMENTED SEVERAL MEASURES TO IMPROVE IT: A. SUCTIONS MACHINES HAVE BEEN INSTALLED IN GREY CLOTH ROLLING PROCESS, DYEING PROCESS AND CUTTING PROCESS. B. THE SUCTION PROCESS WAS ADDED BEFORE PRODUCT'S FINAL FOLDING, AND WORKERS DO IT ACCORDING TO STANDARD OPERATION PROCEDURE REQUIREMENT. C. LINTING TEST METHOD AND ACCEPTABLE CRITERIA WAS STIPULATED TO SEE THE SUCTION RESULTS. (=0.38 G/10 PIECES). D. IN THE FOLDING PROCESS, SUPPLIER USED ONE CLOTH PAD UNDER 100 PIECES SEMI-FINISHED PRODUCTS TO AVOID LINTING STUCK ONTO THE PRODUCTS DURING PRODUCT'S TRANSFER. FROM THE INVESTIGATION, NO ABNORMAL SITUATION OCCURRED DURING PRODUCTION AND THE SAMPLE WAS NOT AVAILABLE; THEREFORE, THE ROOT CAUSE COULD NOT BE DETERMINED. THE COMPLAINT INFORMATION WAS SHARED WITH THE RELEVANT SECTORS FOR THEIR AWARENESS. NO ACTION TAKEN AT THIS TIME, BUT THE SUPPLIER WILL CONTINUE TO MONITOR TRENDS FOR THIS TYPE OF INCIDENT.

Description of Event or Problem · 1

CUSTOMER REPORTS TOO MUCH LINT COMING OFF OF THE TOWELS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
764134 CRANIOTOMY PACK (CNKDC)692 NEUROLOGICAL TRAY OJG MEX03 MEXICO-JUAREZ PRESOURCE SNE69CNKDD UNKNOWN 10888439466560

Patients

Seq Age Sex Outcome Treatment
1