TCM4 COMBI MODULE
Report
- Report Number
- 1523456-2020-00001
- Event Type
- Injury
- Date Received
- July 20, 2020
- Date of Event
- May 13, 2020
- Report Date
- August 14, 2020
- Product Code
- LKD
- Adverse Event
- Yes
- Report Source
- Distributor report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
IT IS NOT POSSIBLE TO CONCLUDE ON AN INVESTIGATION BASED ON THE AVAILABLE INFORMATION IN THIS CASE. THE OPERATOR'S MANUAL DESCRIBES THAT THE SMALLER N20 FIXATION OR ADHESIVE RING SHOULD BE USED FOR NEONATES. THE SMALLER RING MAY BE MORE GENTLE TO THE SKIN DUE TO THE SMALLER ADHESIVE SURFACE. IT WAS RECOMMENDED TO THE CUSTOMER TO TRY THIS SMALLER RING INSTEAD OF THE 32MM RING FOR THIS PATIENT POPULATION. THE CUSTOMER HAS INFORMED THAT THEY WILL CONTINUE TO USE THE LARGER FIXATION RINGS, DESPITE RADIOMETER'S RECOMMENDATION.
IT WAS INFORMED BY THE RADIOMETER REPRESENTATIVE THAT THE HOSPITAL IS USING SENSOR 54 (ITEM NUMBER: 945-736 MANUFACTURED BY RADIOMETER BASEL AG) WITH FIXATION RING 32 MM (ITEM NUMBER: 5601500 MANUFACTURES BY RADIOMETER BASEL AG) IN COMBINATION WITH THE TCM COMBIM MONITOR. MORE INFORMATION HAS BEEN ASKED FROM RADIOMETER TO THE CUSTOMER FOR THE FURTHER INVESTIGATION.
A CUSTOMER REPORTED BRUISING AND REDNESS ON BOTH LEGS BILATERALLY WHERE THE TCM SITES WERE PLACED ON AN ELGAN (EXTREMELY LOW GESTATIONAL AGES NEWBORN) 23 WEEKS OLD NEWBORN. THE DOCTOR FELT THE SKIN WAS SO CONCERNING AND ORDERED FOR A WOUND CONSULTANT.
A CUSTOMER REPORTED BRUISING AND REDNESS ON BOTH LEGS BILATERALLY WHERE THE TCM SITES WERE PLACED ON AN ELGAN (EXTREMELY LOW GESTATIONAL AGES NEWBORN) 23 WEEKS OLD NEWBORN. THE DOCTOR FELT THE SKIN WAS SO CONCERNING AND ORDERED FOR A WOUND CONSULTANT.
A CUSTOMER REPORTED BRUISING AND REDNESS ON BOTH LEGS BILATERALLY WHERE THE TCM SITES WERE PLACED ON AN ELGAN (EXTREMELY LOW GESTATIONAL AGES NEWBORN) 23 WEEKS OLD NEWBORN. THE DOCTOR FELT THE SKIN WAS SO CONCERNING AND ORDERED FOR A WOUND CONSULTANT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 762658 | TCM4 COMBI MODULE | TCM4 COMBI MODULE | LKD | 902-778 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |