FDA Adverse Event Malfunction Summary report: N

PHILIPS MEDICAL SYSTEMS

MDR report key: 1029663 · Received April 15, 2008

Report

Report Number
MW5006271
Event Type
Malfunction
Date Received
April 15, 2008
Date of Event
April 7, 2008
Report Date
April 14, 2008
Manufacturer
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product Code
IZI
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MN, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

BANK OF 4 VIDEO MONITORS FELL FROM THE CEILING SUSPENSION DURING A CATH LAB PROCEDURE. THE MONITOR BANK DID NOT FALL ON THE PT. STAFF MEMBER WAS REPOSITIONING THE MONITORS AT THE TIME OF THE FALL. STAFF PARTIALLY CAUGHT THE SYSTEM AND GUIDED IT TOWARD THE FLOOR. POWER AND VIDEO CABLES KEPT THE MONITORS FROM HITTING THE FLOOR. THIS SYSTEM WEIGHS 200 LB'S OR MORE. IF IT HAD BEEN POSITIONED OVER THE PT AT THE TIME OF THE FALL, IT COULD HAVE CAUSED SERIOUS INJURY. PHILIPS MEDICAL IS AWARE OF THE INCIDENT AND WILL LOOK AT THE ENGINEERING OF THEIR MONITOR SUSPENSION. ALL OTHER ROOMS HAVE BEEN CHECKED TO ENSURE ALL BOLTS ARE TIGHT IN THE MONITOR BANKS. THIS IS AN ASSEMBLY THAT COMES READY MADE FROM THE FACTORY, SO LOCAL INSTALLATION TECHNIQUE IS NOT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PHILIPS MEDICAL SYSTEMS ASSEMBLY FOR MONITORS IZI PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA FD 20 NA

Patients

Seq Age Sex Outcome Treatment