FDA Adverse Event Injury Summary report: N

2.4MM DRILL BIT/QC WITH 65MM STOP

MDR report key: 10295042 · Received July 20, 2020

Report

Report Number
8030965-2020-04983
Event Type
Injury
Date Received
July 20, 2020
Date of Event
May 26, 2020
Report Date
June 29, 2020
Manufacturer
OBERDORF SYNTHES PRODUKTIONS GMBH
Product Code
HTW
UDI-DI
07611819752931
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H6: PART NUMBER:388.394, SYNTHES LOT NUMBER: U242616, SUPPLIER LOT NUMBER: N/A, RELEASE TO WAREHOUSE DATE: MARCH 17, 2016, EXPIRATION DATE: N/A, SUPPLIER: (B)(4). NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUBCOMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. VISUAL INSPECTION: THE EVALUATION OF THE RECEIVED DRILL BIT HAS SHOWN THAT A PIECE OF APPROX. 19MM LENGTH IS BROKEN OFF. THE BROKEN OFF FRAGMENT WAS NOT RETURNED FOR INVESTIGATION. OTHERWISE IS THE DRILL BIT IN GOOD CONDITION, JUST SLIGHT WEAR MARKS ARE AT THE COUPLING PIECE VISIBLE. DIMENSIONAL INSPECTION: CHECKED DIMENSIONS WITH CALIPER 3-03-17585 . OUTER DIAMETER SPECIFICATION 2.4MM 0/-0.05 / MEASURED: 2.37MM = PASS . OVERALL LENGTH (MEASURED WITH A RULER TO DEFINE LENGTH OF MISSING PIECE) SPECIFICATION 205MM +/-1 / MEASURED: 186MM = DAMAGED POST-MANUFACTURING. DRAWING/SPECIFICATION REVIEW: THE DRAWING WAS REVIEWED DURING THE EVALUATION TO VERIFY THE MEASUREMENTS, THE MATERIAL AND THE HARDNESS. THE REVIEW OF THE MANUFACTURING DOCUMENTS HAS SHOWN THAT WITH 440A STAINLESS STEEL THE CORRECT MATERIAL WAS USED AND THAT THE HARDNESS WAS WITH 51.6 HRC WITHIN THE SPECIFICATION OF 50-55 HRC FROM DRAWING . SUMMARY: THE RECEIVED CONDITION OF THE DEVICE IS CONCORDANT WITH THE COMPLAINT DESCRIPTION AND THE COMPLAINT CONDITION IS CONFIRMED. BASED ON THE PROVIDED INFORMATION WE ARE NOT ABLE TO DETERMINE THE EXACT CAUSE OF THIS COMPLAINT. WE ONLY CAN ASSUME THAT A MECHANICAL OVERLOADING SITUATION, FOR EXAMPLE METALLIC CONTACT OR LATERAL STRESS, HAS CAUSED THE BREAKAGE. PLEASE ALSO NOTE; BLUNT DRILL BITS REQUIRE MORE MECHANICAL POWER DURING THE APPLICATION. DO NOT USE SEVERELY DAMAGED INSTRUMENTS, INSTRUMENTS WITH UNRECOGNIZABLE MARKINGS, CORROSION, OR BLUNT CUTTING SURFACES. WE CONCLUDE THAT THE CAUSE OF FAILURE IS NOT DUE TO ANY MANUFACTURING NON-CONFORMANCES. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED AS NO PRODUCT RELATED ISSUE COULD BE DETECTED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H6: THE DEVICE WAS RECEIVED, THE INVESTIGATION IS IN PROGRESS, NO CONCLUSION COULD BE DRAWN AT THE TIME OF FILING THIS REPORT. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION, BUT HAS YET TO BE RECEIVED. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2020, THE SYNAPSE DRILL BIT BROKE AND WAS INSERTED INTO THE PATIENT'S CERVICAL SPINE. PROCEDURE OUTCOME AND PATIENT STATUS ARE UNKNOWN. THIS REPORT IS FOR A 2.4MM DRILL BIT. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
762055 2.4MM DRILL BIT/QC WITH 65MM STOP BIT, DRILL HTW OBERDORF SYNTHES PRODUKTIONS GMBH U242616 07611819752931

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention