FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 10292422 · Received July 17, 2020

Report

Report Number
1644487-2020-00970
Event Type
Malfunction
Date Received
July 17, 2020
Date of Event
June 24, 2020
Report Date
July 16, 2020
Manufacturer
LIVANOVA USA, INC.
Product Code
LYJ
UDI-DI
05425025750443
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

INTERNAL REFERENCE NUMBER: (B)(4). VOLUNTARY MEDICAL DEVICE CORRECTION (REMEDIAL ACTION) WAS INITIATED BY THE MANUFACTURER ON 2/4/2020, AND THE PHYSICIAN WAS PROVIDED A NOTIFICATION LETTER WITH RECOMMENDED ACTIONS. SUSPECT DEVICE - SOFTWARE VERSION IS (B)(4).

Description of Event or Problem · 1

LOW OUTPUT CURRENT WAS SEEN AFTER TURNING THE PATIENT'S DEVICE ON AFTER IT WAS DISABLED FOR AN MRI. THE TABLET WAS RESTARTED AND DIAGNOSTICS WERE RUN AND THE RESULTS WERE WITHIN NORMAL LIMITS. THE GENERATOR WAS INTERROGATED WITH (B)(4) SOFTWARE. INTERNAL TESTING AND DATA REVIEW IDENTIFIED THAT FALSE LOW OUTPUT CURRENT MESSAGES CAN OCCUR WHEN M3000 (B)(4) SOFTWARE PROGRAMS A M103-106 GENERATOR AFTER A SPECIFIC PROGRAMMING SEQUENCE OCCURS. WHEN M3000 (B)(4) TABLETS INITIATE A PROGRAMMING SESSION WITH A M103-106 GENERATOR, WHERE ITS OUTPUT CURRENT IS PROGRAMMED OFF (0 MA), AND THEN THE OUTPUT CURRENT IS PROGRAMMED BACK ON TO >0 MA, A LOW OUTPUT CURRENT MESSAGE WOULD BE SEEN WHEN AN IN-SESSION INTERROGATION IS PERFORMED. THIS WOULD PERSIST UPON MULTIPLE IN-SESSION INTERROGATIONS AND IF DIAGNOSTICS WEREN'T PERFORMED DURING THE SESSION, SHOW ON THE SESSION REPORT. RUNNING SYSTEM DIAGNOSTICS OR ENDING AND RESTARTING THE SESSION WILL CONFIRM FUNCTIONALITY OF THE DEVICE, AND RESOLVE THE ¿OUTPUT CURRENT LOW¿ ERROR MESSAGE. THE LOW OUTPUT CURRENT MESSAGES IN THESE CASES ARE FALSE AND DO NOT IMPACT GENERATOR FUNCTION. NO FURTHER RELEVANT INFORMATION HAS BEEN RECEIVED TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
755330 PROGRAMMING SOFTWARE PROGRAMMING SOFTWARE LYJ LIVANOVA USA, INC. MODEL 3000 05425025750443

Patients

Seq Age Sex Outcome Treatment
1 9 YR