FDA Adverse Event Injury Summary report: N

PERMOBIL C500

MDR report key: 10291273 · Received July 17, 2020

Report

Report Number
1221084-2020-00034
Event Type
Injury
Date Received
July 17, 2020
Date of Event
June 17, 2020
Report Date
July 17, 2020
Manufacturer
PERMOBIL AB (PAB)
Product Code
ITI
PMA / PMN Number
K991658
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

REPORTS RECEIVED BY PERMOBIL BELGIUM CLAIM WHILE THE END-USER WAS DRIVING THEIR DEVICE DOWN A SIDEWALK AT SPEED, THE DEVICE SUDDENLY STOPPED CAUSING THE END-USER TO LOSE THEIR POSITIONING AND FALL OUT OF THE SEATING TO THE GROUND. REPORTS INDICATE THE FALL RESULTED IN JURIES CONSISTING OF FRACTURED TEETH AND LACERATION TO THE CHIN REQUIRING STITCHES. INSPECTION OF THE DEVICE WAS PERFORMED BY PERMOBIL TECHNICIANS WHERE FINDINGS SHOWN DAMAGES TO ONE OF THE ELECTRICAL CABLES FOR THE RIGHT MOTOR BRAKE. THIS DAMAGE WOULD CAUSE AN OPEN CIRCUIT WHICH RESULTS IN LOSS OF CONTINUITY, INTERRUPTING THE POWER TO THE ELECTRIC BRAKE. THIS LOSS OF POWER TO THE BRAKE, IF IT WERE TO RECUR WHILE DRIVING, WOULD CAUSE THE MOTOR/GEARBOX TO STOP ROTATING. THIS WOULD EXPLAIN THE SUDDEN STOP AS DESCRIBED BY THE END-USER. INSPECTION SHOWN A PLASTIC END-CAP WHICH ATTACHES TO THE END OF THE UNITRACK SEAT RAIL HAVING FALLEN OFF AND BECOME WEDGED BETWEEN THE MOTOR/BRAKE AND CHASSIS. IT IS UNCLEAR AS TO HOW LONG THIS END CAP HAD BEEN IN THIS POSITION, BUT EVALUATION RESULTS SHOWN THIS WEDGED COMPONENT HAVING DAMAGED ONE OF THE POWER WIRES LEADING TO THE RIGHT BRAKE SOLENOID. REVIEW OF COMPLAINT DATA INDICATES THIS TO BE AN ISOLATED EVENT WITH NO RECORDS TO INDICATE THIS FAILURE MODE HAVING BEEN PREVIOUSLY REPORTED. IT WAS NOTED THAT THE POSITIONING BELT WHICH WAS PROVIDED WHEN DEVICE WAS ORIGINALLY DISTRIBUTED HAD BEEN REMOVED AT SOME POINT SINCE DISTRIBUTION AND WAS NOT BEING USED AT THE TIME OF THE INCIDENT. DEVICE REMAINS AT PERMOBIL BELGIUM'S SERVICE CENTER UNTIL A DETERMINATION IS MADE BETWEEN HEALTHCARE FACILITY, INSURANCE PROVIDER AND END USER AS TO IF THIS 4 YEAR OLD DEVICE WILL BE REPAIRED OR REPLACED. THE DHR WAS REVIEWED AND THE DEVICE MET SPECIFICATION PRIOR TO DISTRIBUTION.

Description of Event or Problem · 1

RECEIVED REPORT CLAIMING WHILE WAS USER WAS DRIVING BESIDES HIS SON WHO WAS ON HIS BIKE. USER WAS DRIVING MAXIMUM SPEED (10 KM/U) WHEN SUDDENLY THE DRIVE MOTOR STOPPED. REPORTS CLAIM END-USER LOST POSITIONING ALLOWING THEM TO FALL OUT OF THE SEATING TO THE GROUND RESULTING IN INJURY REQUIRING MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
754922 PERMOBIL C500 POWERED WHEELCHAIR ITI PERMOBIL AB (PAB) C500 N/A

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization