FDA Adverse Event Malfunction Summary report: N

ZOLL IVTM THERMOGARD XP

MDR report key: 10287491 · Received July 17, 2020

Report

Report Number
3010617000-2020-00682
Event Type
Malfunction
Date Received
July 17, 2020
Date of Event
June 22, 2020
Report Date
September 26, 2020
Manufacturer
ZOLL CIRCULATION
Product Code
NCX
UDI-DI
00849111075022
PMA / PMN Number
K072234
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE INITIAL CONSOLE EVALUATION WAS PERFORMED AZM SERVICE DEPOT AND DURING THE TESTING, THE CUSTOMER REPORTED COMPLAINT OF BLANK DISPLAY PANEL SCREEN AND A BEEPING TONE WAS CONFIRMED, HOWEVER, THE ISSUE WAS OBSERVED INTERMITTENTLY. THE DISPLAY HEAD WAS RETURNED TO ZOLL AND UPON EVALUATION, THE REPORTED ISSUE COULD NOT BE DUPLICATED DURING THE FUNCTIONAL TESTING. THE PROBABLE CAUSE OF THE REPORTED ISSUE WAS LIKELY DUE TO THE LOOSE CONNECTION OF UMBILICAL CABLE FROM PRINTED CIRCUIT ASSEMBLY (PCA) TO THE DISPLAY HEAD MONITOR. THE THERMOGARD CONSOLE WAS MANUFACTURED IN SEPTEMBER 2013 AND IS ALMOST 7 YEARS OLD, PAST THE EXPECTED SERVICE LIFE OF 5 YEARS. DURING THE VISUAL INSPECTION, THERE WAS NO PHYSICAL DAMAGE OBSERVED ON THE DISPLAY HEAD. THE EVENT LOG REVIEW DONE BY AZM SERVICE DEPOT SHOWED NO SIGNIFICANT DISCREPANCIES. THE DISPLAY HEAD WAS CONNECTED TO THE GOOD KNOWN IVTM THERMOGARD CONSOLE AND PASSED THE INITIAL FUNCTIONAL TESTING AND A 5 DAYS BURN-IN TEST WITH NO ISSUES. THERE WAS NO DEVICE MALFUNCTION ON THE DISPLAY HEAD MONITOR, AND THE MONITOR FUNCTIONED AS INTENDED. ALSO, THE CONSOLE SN (B)(6) WAS TESTED IN THE FIELD WITH ANOTHER DISPLAY HEAD AND PASSED FUNCTIONAL TESTING WITHOUT ANY ERROR OR FAULT.

Additional Manufacturer Narrative · 0

THE INITIAL CONSOLE EVALUATION WAS PERFORMED AZM SERVICE DEPOT AND DURING THE TESTING, THE CUSTOMER REPORTED COMPLAINT OF BLANK DISPLAY PANEL SCREEN AND A BEEPING TONE WAS CONFIRMED, HOWEVER, THE ISSUE WAS OBSERVED INTERMITTENTLY. THE DISPLAY HEAD WAS RETURNED TO ZOLL AND UPON EVALUATION, THE REPORTED ISSUE COULD NOT BE DUPLICATED DURING THE FUNCTIONAL TESTING. THE PROBABLE CAUSE OF THE REPORTED ISSUE WAS LIKELY DUE TO THE LOOSE CONNECTION OF UMBILICAL CABLE FROM PRINTED CIRCUIT ASSEMBLY (PCA) TO THE DISPLAY HEAD MONITOR. THE THERMOGARD CONSOLE WAS MANUFACTURED IN SEPTEMBER 2013 AND IS ALMOST 7 YEARS OLD, PAST THE EXPECTED SERVICE LIFE OF 5 YEARS. DURING THE VISUAL INSPECTION, THERE WAS NO PHYSICAL DAMAGE OBSERVED ON THE DISPLAY HEAD. THE EVENT LOG REVIEW DONE BY AZM SERVICE DEPOT SHOWED NO SIGNIFICANT DISCREPANCIES. THE DISPLAY HEAD WAS CONNECTED TO THE GOOD KNOWN IVTM THERMOGARD CONSOLE AND PASSED THE INITIAL FUNCTIONAL TESTING AND A 5 DAYS BURN-IN TEST WITH NO ISSUES. THERE WAS NO DEVICE MALFUNCTION ON THE DISPLAY HEAD MONITOR, AND THE MONITOR FUNCTIONED AS INTENDED. AS A PRECAUTIONARY MEASURE, THE LCD INVERTER IN THE DISPLAY HEAD WAS REPLACED. ALSO, THE CONSOLE SN (B)(6) WAS TESTED IN THE FIELD WITH ANOTHER DISPLAY HEAD AND PASSED FUNCTIONAL TESTING WITHOUT ANY ERROR OR FAULT.

Additional Manufacturer Narrative · 1

THE ZOLL HAS RECEIVED THE DEVICE FOR INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE FILED WHEN THE INVESTIGATION HAS BEEN COMPLETED.

Description of Event or Problem · 1

DURING PREVENTIVE MAINTENANCE, THE THERMOGARD XP SYSTEM (SN (B)(4)) PRODUCED A BEEPING TONE AND A BLANK DISPLAY PANEL SCREEN UPON POWER ON. NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
751675 ZOLL IVTM THERMOGARD XP THERMAL REGULATING SYSTEM NCX ZOLL CIRCULATION 8700-0650 00849111075022

Patients

Seq Age Sex Outcome Treatment
1