ZOLL IVTM THERMOGARD XP
Report
- Report Number
- 3010617000-2020-00682
- Event Type
- Malfunction
- Date Received
- July 17, 2020
- Date of Event
- June 22, 2020
- Report Date
- September 26, 2020
- Manufacturer
- ZOLL CIRCULATION
- Product Code
- NCX
- UDI-DI
- 00849111075022
- PMA / PMN Number
- K072234
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
THE INITIAL CONSOLE EVALUATION WAS PERFORMED AZM SERVICE DEPOT AND DURING THE TESTING, THE CUSTOMER REPORTED COMPLAINT OF BLANK DISPLAY PANEL SCREEN AND A BEEPING TONE WAS CONFIRMED, HOWEVER, THE ISSUE WAS OBSERVED INTERMITTENTLY. THE DISPLAY HEAD WAS RETURNED TO ZOLL AND UPON EVALUATION, THE REPORTED ISSUE COULD NOT BE DUPLICATED DURING THE FUNCTIONAL TESTING. THE PROBABLE CAUSE OF THE REPORTED ISSUE WAS LIKELY DUE TO THE LOOSE CONNECTION OF UMBILICAL CABLE FROM PRINTED CIRCUIT ASSEMBLY (PCA) TO THE DISPLAY HEAD MONITOR. THE THERMOGARD CONSOLE WAS MANUFACTURED IN SEPTEMBER 2013 AND IS ALMOST 7 YEARS OLD, PAST THE EXPECTED SERVICE LIFE OF 5 YEARS. DURING THE VISUAL INSPECTION, THERE WAS NO PHYSICAL DAMAGE OBSERVED ON THE DISPLAY HEAD. THE EVENT LOG REVIEW DONE BY AZM SERVICE DEPOT SHOWED NO SIGNIFICANT DISCREPANCIES. THE DISPLAY HEAD WAS CONNECTED TO THE GOOD KNOWN IVTM THERMOGARD CONSOLE AND PASSED THE INITIAL FUNCTIONAL TESTING AND A 5 DAYS BURN-IN TEST WITH NO ISSUES. THERE WAS NO DEVICE MALFUNCTION ON THE DISPLAY HEAD MONITOR, AND THE MONITOR FUNCTIONED AS INTENDED. ALSO, THE CONSOLE SN (B)(6) WAS TESTED IN THE FIELD WITH ANOTHER DISPLAY HEAD AND PASSED FUNCTIONAL TESTING WITHOUT ANY ERROR OR FAULT.
THE INITIAL CONSOLE EVALUATION WAS PERFORMED AZM SERVICE DEPOT AND DURING THE TESTING, THE CUSTOMER REPORTED COMPLAINT OF BLANK DISPLAY PANEL SCREEN AND A BEEPING TONE WAS CONFIRMED, HOWEVER, THE ISSUE WAS OBSERVED INTERMITTENTLY. THE DISPLAY HEAD WAS RETURNED TO ZOLL AND UPON EVALUATION, THE REPORTED ISSUE COULD NOT BE DUPLICATED DURING THE FUNCTIONAL TESTING. THE PROBABLE CAUSE OF THE REPORTED ISSUE WAS LIKELY DUE TO THE LOOSE CONNECTION OF UMBILICAL CABLE FROM PRINTED CIRCUIT ASSEMBLY (PCA) TO THE DISPLAY HEAD MONITOR. THE THERMOGARD CONSOLE WAS MANUFACTURED IN SEPTEMBER 2013 AND IS ALMOST 7 YEARS OLD, PAST THE EXPECTED SERVICE LIFE OF 5 YEARS. DURING THE VISUAL INSPECTION, THERE WAS NO PHYSICAL DAMAGE OBSERVED ON THE DISPLAY HEAD. THE EVENT LOG REVIEW DONE BY AZM SERVICE DEPOT SHOWED NO SIGNIFICANT DISCREPANCIES. THE DISPLAY HEAD WAS CONNECTED TO THE GOOD KNOWN IVTM THERMOGARD CONSOLE AND PASSED THE INITIAL FUNCTIONAL TESTING AND A 5 DAYS BURN-IN TEST WITH NO ISSUES. THERE WAS NO DEVICE MALFUNCTION ON THE DISPLAY HEAD MONITOR, AND THE MONITOR FUNCTIONED AS INTENDED. AS A PRECAUTIONARY MEASURE, THE LCD INVERTER IN THE DISPLAY HEAD WAS REPLACED. ALSO, THE CONSOLE SN (B)(6) WAS TESTED IN THE FIELD WITH ANOTHER DISPLAY HEAD AND PASSED FUNCTIONAL TESTING WITHOUT ANY ERROR OR FAULT.
THE ZOLL HAS RECEIVED THE DEVICE FOR INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE FILED WHEN THE INVESTIGATION HAS BEEN COMPLETED.
DURING PREVENTIVE MAINTENANCE, THE THERMOGARD XP SYSTEM (SN (B)(4)) PRODUCED A BEEPING TONE AND A BLANK DISPLAY PANEL SCREEN UPON POWER ON. NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 751675 | ZOLL IVTM THERMOGARD XP | THERMAL REGULATING SYSTEM | NCX | ZOLL CIRCULATION | 8700-0650 | 00849111075022 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |