FDA Adverse Event Malfunction Summary report: N

SYNCHRON LX I 725 CLINICAL SYSTEM

MDR report key: 1028600 · Received April 10, 2008

Report

Report Number
2122870-2008-00117
Event Type
Malfunction
Date Received
April 10, 2008
Date of Event
March 25, 2008
Report Date
April 10, 2008
Manufacturer
BECKMAN COULTER, INC.
Product Code
MMI
PMA / PMN Number
K023049
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
NO INFORMATION

Narratives

Additional Manufacturer Narrative · 1

QC AND SYSTEM CHECK WERE WITHIN SPECS BEFORE THIS EVENT. THE SPECIMENS WERE COLLECTED IN LITHIUM HEPARIN TUBES WITH GEL AND CENTRIFUGED AT 3,500 RPM FOR 5 MINS. BD PRODUCT INSERT RECOMMENDS SAMPLES TO BE CENTRIFUGED FOR 10 MINUTES. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S LAB: THE FSE INSPECTED THE INSTRUMENT AND DID NOT FIND ANY HARDWARE RELATED ISSUES THAT MAY HAVE CONTRIBUTED TO THIS EVENT. ALL VERIFICATION TESTING PASSED WITHIN SPECS. PRE-ANALYTICAL SAMPLE HANDLING WAS DISCUSSED WITH THE CUSTOMER, AND THE FSE GAVE THEM LITERATURE TO DISTRIBUTE TO ALL OF STAFF AND THE ER FOR FUTURE REFERENCE. PRE-ANALYTICAL SAMPLE HANDLING MAY HAVE CONTRIBUTED TO THIS EVENT; HOWEVER, A CLEAR ROOT CAUSE FOR THIS EVENT HAS NOT BEEN DETERMINED TO DATE. A MALFUNCTION WILL BE ASSUMED FOR THE PURPOSE OF THIS REPORT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER, INC (BCI) REGARDING AN ERRONEOUSLY ELEVATED TROPONIN (ACCU TNI) RESULT THAT WAS GENERATED BY THE SYNCHRON LX I 725 INSTRUMENT FOR A SINGLE PT SAMPLE. A PT SAMPLE WAS TESTED FOR ACCU TNI AND A RESULT OF 14.0NG/ML WAS OBTAINED. BECAUSE OF A SIMILAR EVENT IN THE PAST, CUSTOMER COLLECTED A FRESH SAMPLE FROM THE PT, ENSURING THE SAMPLE INTEGRITY. THE NEW SAMPLE GAVE A RESULT OF 0.03NG/ML AND WAS REPORTED OUT OF THE LAB. NO REPORTS OF DEATH, INJURY OR CHANGE TO PT TREATMENT HAVE BEEN RECEIVED IN CONNECTION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHRON LX I 725 CLINICAL SYSTEM DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER MMI BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA