STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2020-10891
- Event Type
- Injury
- Date Received
- July 16, 2020
- Date of Event
- June 16, 2020
- Report Date
- December 23, 2020
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- UDI-DI
- 10888628000230
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B.3, B.5, D.7.
DEVICE EVALUATION: VISUAL ANALYSIS OF THE PHOTOS IDENTIFIED YELLOW PARTICLE MATERIAL ON THE SHELL; CREASE AND WEAR ABRASION. DEVICE PATCH WITH LOT NUMBER 3152539 AND CATALOG NUMBER 68MP-360. DEVICE ANALYSIS PERFORMED THROUGH PHOTOGRAPHS, DUE TO THE IMPOSSIBILITY TO PERFORM MICROSCOPIC ANALYSIS AS IT WAS NOT POSSIBLE TO DETERMINE THE MOST LIKELY FAILURE MODE.
THE DEVICE HAS BEEN EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. THE DEVICE HAS BEEN EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN INITIATED. IF ANY NEW, CHANGED OR CORRECTED INFORMATION IS NOTED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE DEFLATION. THE DEVICE REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 745697 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 68-360 | 3152539 | 10888628000230 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR | Required Intervention |