T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY
Report
- Report Number
- 3013756811-2020-73511
- Event Type
- Malfunction
- Date Received
- July 16, 2020
- Date of Event
- June 10, 2020
- Report Date
- July 16, 2020
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- QFG
- UDI-DI
- 00852162004781
- PMA / PMN Number
- DEN180058
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
THE T:SLIM X2 WITH CIQ TECHNOLOGY USER GUIDE STATES: DO NOT USE ANY OTHER INSULIN WITH YOUR SYSTEM OTHER THAN U-100 HUMALOG OR U-100 NOVOLOG. ONLY HUMALOG AND NOVOLOG HAVE BEEN TESTED AND FOUND TO BE COMPATIBLE FOR USE IN THE SYSTEM. USE OF INSULIN WITH LESSER OR GREATER CONCENTRATION CAN RESULT IN UNDER DELIVERY OR OVER DELIVERY OF INSULIN. THIS CAN CAUSE VERY HIGH OR A VERY LOW BLOOD GLUCOSE. NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT INTERMITTENT OCCLUSION ALARMS OCCURRED. CUSTOMER REPORTED USING APIDRA INSULIN. TANDEM CUSTOMER TECHNICAL SUPPORT INFORMED CUSTOMER THAT APIDRA IS OFF LABEL PER THE USER GUIDE. CUSTOMER CHANGED PUMP SUPPLIES TO ADDRESS THE OCCLUSION ALARMS AND SUCCESSFULLY RESUMED INSULIN. CUSTOMER'S BLOOD GLUCOSE RANGED FROM 200-326 MG/DL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 745387 | T:SLIM X2 INSULIN PUMP WITH INTEROPERABLE TECHNOLOGY | ALTERNATE CONTROLLER ENABLED INFUSION PUMP | QFG | TANDEM DIABETES CARE | 1000096 | 00852162004781 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR |