FDA Adverse Event Malfunction Summary report: N

DATEX OHMEDA AEDSTIVA/5

MDR report key: 1028491 · Received April 10, 2008

Report

Report Number
MW5006228
Event Type
Malfunction
Date Received
April 10, 2008
Date of Event
April 10, 2008
Report Date
April 10, 2008
Manufacturer
DATEX OHMED
Product Code
CAD
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NV, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

DATEX OHMEDA AESTIVA/5 ANESTHESIA MACHINE WITH PSVPRO VENTILATOR; VENTILATOR TURNED ON VT600 RATE 10. NOTICED AFTER APPROX 2 MINUTES THAT VENTILATION HAD NOT INITIATED. ON SCREEN VENTILATOR NOTIFICATION "INSPIRATORY FLOW SENSOR FAILURE" PATIENT'S O2 SAT REMAINED 100% -BRIEF PERIOD/APNEIC VENTILATION-. IN THE EVENT OF A MECHANICAL VENTILATOR FAILURE, THE VENTILATOR ALARM SHOULD ANNUNCIATE; IT DID NOT...REPLACED FLOW SENSOR MODULE; NORMAL FUNCTION RE-INITIATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DATEX OHMEDA AEDSTIVA/5 ANESTHESIA MACHINE/VENTILATOR CAD DATEX OHMED AESTIVA 7900CON

Patients

Seq Age Sex Outcome Treatment
1 Other