FDA Adverse Event
Malfunction
Summary report: N
DATEX OHMEDA AEDSTIVA/5
MDR report key: 1028491
·
Received April 10, 2008
Report
- Report Number
- MW5006228
- Event Type
- Malfunction
- Date Received
- April 10, 2008
- Date of Event
- April 10, 2008
- Report Date
- April 10, 2008
- Manufacturer
- DATEX OHMED
- Product Code
- CAD
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- NV, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
DATEX OHMEDA AESTIVA/5 ANESTHESIA MACHINE WITH PSVPRO VENTILATOR; VENTILATOR TURNED ON VT600 RATE 10. NOTICED AFTER APPROX 2 MINUTES THAT VENTILATION HAD NOT INITIATED. ON SCREEN VENTILATOR NOTIFICATION "INSPIRATORY FLOW SENSOR FAILURE" PATIENT'S O2 SAT REMAINED 100% -BRIEF PERIOD/APNEIC VENTILATION-. IN THE EVENT OF A MECHANICAL VENTILATOR FAILURE, THE VENTILATOR ALARM SHOULD ANNUNCIATE; IT DID NOT...REPLACED FLOW SENSOR MODULE; NORMAL FUNCTION RE-INITIATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DATEX OHMEDA AEDSTIVA/5 | ANESTHESIA MACHINE/VENTILATOR | CAD | DATEX OHMED | AESTIVA 7900CON |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |