LIFEPAK 9 DEFIBRILLATOR/MONITOR
Report
- Report Number
- 3015876-2008-00316
- Event Type
- Malfunction
- Date Received
- April 10, 2008
- Date of Event
- January 10, 2008
- Report Date
- March 11, 2008
- Manufacturer
- PHYSIO-CONTROL, INC.
- Product Code
- LDD
- PMA / PMN Number
- K881153
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
PHYSIO-CONTROL REPLACED IC CHIP, DESIGNATOR U28, ON THE MAIN PCB ASSEMBLY AND THEN OBSERVED PROPER DEVICE OPERATION THROUGH FUNCTIONAL AND PERFORMANCE TESTING. THE DEVICE WAS RETURNED TO THE CUSTOMER FOR USE. DEVICE DATA STORAGE, INCLUDING DEFIBRILLATOR CALIBRATION CONSTANTS, IS STORED IN A BATTERY-BACKED RAM CHIP, DESIGNATOR U28, ON THE MAIN PCB ASSEMBLY. IC CHIP U28 CONTAINS A LITHIUM ENERGY CELL BATTERY THAT CAN MAINTAIN THE DEVICE DATA FOR A MINIMUM OF 5 YEARS CONTINUOUS DUTY. A RECENT FAILURE ANALYSIS HAS DETERMINED THAT A SEVERELY DEPLETED ENERGY CELL CAN RESULT IN A FAILURE OF THE DEFIBRILLATOR TO CHARGE COMPLETELY AND/OR A FAILURE TO DISCHARGE.
FOUND DURING REGULAR INSPECTION AND TESTING. PHYSIO-CONTROL OBSERVED THAT THE DEVICE DISPLAYS A SERVICE INDICATOR WHEN POWERING UP, AN "ENERGY FAULT" WARNING MESSAGE WHEN THE CHARGE BUTTON WAS PRESSED, AND A FAULT CODE 22 WAS LOGGED INTO THE DEVICE ERROR LOG.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEPAK 9 DEFIBRILLATOR/MONITOR | LDD | PHYSIO-CONTROL, INC. | 9 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |