CABLE SIEMENS 10P
Report
- Report Number
- 1226348-2020-00169
- Event Type
- Malfunction
- Date Received
- July 15, 2020
- Report Date
- March 12, 2020
- Manufacturer
- INTEGRA LIFESCIENCES MANSFIELD
- Product Code
- GWM
- PMA / PMN Number
- K945585
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPDATED FIELDS: D10, G4, G7, H2, H3, H4, H6, H10 DEVICE IDENTIFIER # (B)(4). CABLE WAS RETURNED FOR EVALUATION. DHR - THE REVIEW REVEALED NO ANOMALIES FAILURE ANALYSIS WAS COMPLETED, AND THE UNIT PASSED ALL FUNCTIONAL TESTING. THE ROOT CAUSE COULD THEREFORE NOT BE DETERMINED. THE AS FOUND CONDITION DID REVEAL SOME KINKING AND DAMAGED THREADS. BETWEEN 05NOV2019 AND 30JUN2020, APPROXIMATELY 2,200 MDRS SUBMITTED ELECTRONICALLY BY INTEGRA LIFESCIENCES VIA TRACKWISE, INTEGRA'S COMPLAINT HANDLING SYSTEM, WERE NOT RECEIVED BY CDRH DUE TO A COMPUTER SYSTEM ISSUE. WITHIN THIS TIME PERIOD, AN ERROR WITH INTEGRA'S MIDDLEWARE, WHICH FACILITATES COMMUNICATIONS BETWEEN TRACKWISE AND THE FDA SYSTEM, CAUSED THE COMPLAINT RECORDS TO CLOSE AND INDICATE WE HAD RECEIVED AN ACKNOWLEDGEMENT 3 FROM THE FDA WHEN WE HAD NOT. INTEGRA INTERPRETED THE ACKNOWLEDGEMENT AS A SUCCESSFUL SUBMISSION; HOWEVER, SUBSEQUENT INVESTIGATION REVEALED THE ACKNOWLEDGEMENT 3 RECEIVED WAS FROM OUR MIDDLEWARE AND NOT FROM THE FDA (THESE ACKNOWLEDGEMENTS HAVE BEEN RETAINED AS PART OF THE DOCUMENTATION OF THE MDR). THE MALFUNCTION WAS RELATED TO THE RELOCATION OF THE TRACKWISE APPLICATION TO A NEW DATA CENTER DURING THE TRANSITION OF INTEGRA'S CORPORATE HEADQUARTERS FROM PLAINSBORO, NJ TO PRINCETON, NJ. PREVIOUSLY, INTEGRA HAD BEEN SUCCESSFULLY RECEIVING ACKNOWLEDGEMENTS 1, 2, AND 3 FROM THE FDA, AND OUR RECORDS REFLECT THESE ACKNOWLEDGEMENTS, INCLUDING THE DATE AND TIME STAMPS. CAPA PR 229048 AND NC 20-011 HAVE BEEN OPENED BY INTEGRA TO FURTHER INVESTIGATE THE NONCONFORMANCE AND DEVELOP A CORRECTIVE ACTION PLAN. THE MIDDLEWARE ERROR HAS BEEN CORRECTED, AND INTEGRA HAS FILED MDRS SINCE THE CORRECTION AND VERIFIED THAT THE APPROPRIATE ACKNOWLEDGEMENTS HAVE BEEN RECEIVED FROM THE FDA. INTEGRA IS RESUBMITTING ALL IMPACTED MDR REPORTS FOR THE TIME PERIOD 05NOV2019 THROUGH 30JUN2020. INTEGRA LIFESCIENCES CONTACTED (B)(4) OFFICE OF PRODUCT EVALUATION AND QUALITY ON JULY 8-9, 2020 TO REPORT THESE ISSUES REGARDING MDR REPORTS.
ATTEMPTS ARE BEING MADE TO OBTAIN ADDITIONAL INFORMATION. UPON COMPLETION OF THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.
2 OF 3 REPORTS. OTHER MFG REPORT NUMBERS: 1226348-2020-00168. 1226348-2020-00170. A FACILITY REPORTED AN ICP CABLE MALFUNCTION. A CERELINK SENSOR WAS IMPLANTED. SHORTLY THEREAFTER, THE ICP EXPRESS MONITOR SHOWED -99MMHG. THE CUSTOMER THINKS IT WAS THE CERELINK SENSOR BUT IT COULD ALSO BE THE CONNECTING CABLE OR THE MONITOR. WHEN THE SALES REP RECEIVED THESE ITEMS, SHE WAS ALSO GIVEN 3 SIEMENS INTERFACE CABLES. THIS COMPLAINT IS REGISTERING THE SECOND OF THESE THREE CABLES. MONITORING WAS DISCONTINUED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 739310 | CABLE SIEMENS 10P | ICP EXPRESS - CABLES | GWM | INTEGRA LIFESCIENCES MANSFIELD | Y7487 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |