FDA Adverse Event Injury Summary report: N

SPECIAL DEVICE UNITIZED PROGRA

MDR report key: 10276975 · Received July 15, 2020

Report

Report Number
1226348-2019-00605
Event Type
Injury
Date Received
July 15, 2020
Date of Event
November 26, 2019
Report Date
November 26, 2019
Manufacturer
INTEGRA LIFESCIENCES SWITZERLAND SAR
Product Code
JXG
PMA / PMN Number
NA
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

PRODUCT WAS RECEIVED FOR EVALUATION: DHR - LOT 127783, CONFORMED TO THE SPECIFICATIONS WHEN RELEASED TO STOCK FAILURE ANALYSIS - THE VALVE PASSED THE TESTS FOR OCCLUSIONS, PROGRAMMING, LEAKS, REFLUX, SIPHON GUARD AND PRESSURE. NO ROOT CAUSE COULD BE DETERMINED, AS THE TECHNICIAN WAS UNABLE TO CONFIRM THE PROBLEM REPORTED BY THE CUSTOMER. BETWEEN 05NOV2019 AND 30JUN2020, APPROXIMATELY 2,200 MDRS SUBMITTED ELECTRONICALLY BY INTEGRA LIFESCIENCES VIA TRACKWISE, INTEGRA'S COMPLAINT HANDLING SYSTEM, WERE NOT RECEIVED BY CDRH DUE TO A COMPUTER SYSTEM ISSUE. WITHIN THIS TIME PERIOD, AN ERROR WITH INTEGRA'S MIDDLEWARE, WHICH FACILITATES COMMUNICATIONS BETWEEN TRACKWISE AND THE FDA SYSTEM, CAUSED THE COMPLAINT RECORDS TO CLOSE AND INDICATE WE HAD RECEIVED AN ACKNOWLEDGEMENT 3 FROM THE FDA WHEN WE HAD NOT. INTEGRA INTERPRETED THE ACKNOWLEDGEMENT AS A SUCCESSFUL SUBMISSION; HOWEVER, SUBSEQUENT INVESTIGATION REVEALED THE ACKNOWLEDGEMENT 3 RECEIVED WAS FROM OUR MIDDLEWARE AND NOT FROM THE FDA (THESE ACKNOWLEDGEMENTS HAVE BEEN RETAINED AS PART OF THE DOCUMENTATION OF THE MDR). THE MALFUNCTION WAS RELATED TO THE RELOCATION OF THE TRACKWISE APPLICATION TO A NEW DATA CENTER DURING THE TRANSITION OF INTEGRA'S CORPORATE HEADQUARTERS FROM PLAINSBORO, NJ TO PRINCETON, NJ. PREVIOUSLY, INTEGRA HAD BEEN SUCCESSFULLY RECEIVING ACKNOWLEDGEMENTS 1, 2, AND 3 FROM THE FDA, AND OUR RECORDS REFLECT THESE ACKNOWLEDGEMENTS, INCLUDING THE DATE AND TIME STAMPS. CAPA PR 229048 AND NC 20-011 HAVE BEEN OPENED BY INTEGRA TO FURTHER INVESTIGATE THE NONCONFORMANCE AND DEVELOP A CORRECTIVE ACTION PLAN. THE MIDDLEWARE ERROR HAS BEEN CORRECTED, AND INTEGRA HAS FILED MDRS SINCE THE CORRECTION AND VERIFIED THAT THE APPROPRIATE ACKNOWLEDGEMENTS HAVE BEEN RECEIVED FROM THE FDA. INTEGRA IS RESUBMITTING ALL IMPACTED MDR REPORTS FOR THE TIME PERIOD 05NOV2019 THROUGH 30JUN2020. INTEGRA LIFESCIENCES CONTACTED (B)(6) , DIRECTOR OF REGULATORY PROGRAMS, OFFICE OF PRODUCT EVALUATION AND QUALITY AND (B)(6) , ASSISTANT DIRECTOR, MDR TEAM, OFFICE OF PRODUCT EVALUATION AND QUALITY ON JULY 8-9, 2020 TO REPORT THESE ISSUES REGARDING MDR REPORTS.

Description of Event or Problem · 0

N/A.

Additional Manufacturer Narrative · 1

TO DATE THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE VALVE WAS IMPLANTED TO THE PATIENT VIA L-P SHUNT ON 2016 (EXACT DATE IS UNKNOWN) DUE TO IDIOPATHIC NORMAL-PRESSURE HYDROCEPHALUS. THE INITIAL SETTING IS UNKNOWN. THE VALVE WAS USED WITH THE SILASCON® LUMBAR CATHETER (MANUFACTURED BY KANEKA). DURING THE FOLLOW-UP, VENTRICLE ENLARGEMENT WAS OBSERVED, AND OCCLUSION WAS SUSPECTED. THE VALVE WAS REPLACED WITH A NEW L-P SHUNT VALVE (CERTAS JKTLP8806) ON (B)(6) 2019. THE SETTING WAS UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
740868 SPECIAL DEVICE UNITIZED PROGRA CHPV WITH SG JXG INTEGRA LIFESCIENCES SWITZERLAND SAR

Patients

Seq Age Sex Outcome Treatment
1