FDA Adverse Event Malfunction Summary report: N

IMPL TAPERED SP 4.8MM 10M M OCTAGON

MDR report key: 10276462 · Received July 15, 2020

Report

Report Number
0002023141-2020-01052
Event Type
Malfunction
Date Received
July 15, 2020
Date of Event
May 22, 2020
Report Date
December 15, 2020
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024017566
PMA / PMN Number
K082639
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). A IMPL TAPERED SP 4.8MM 10M M OCTAGON (SPWB10) WAS RETURNED FOR INVESTIGATION. VISUAL INSPECTION OF THE AS RETURNED PRODUCT IDENTIFIED SLIGHT BONE ON THREADS AND DAMAGED INTERNAL FEATURE/THREADS. DEVICE HISTORY RECORD (DHR) REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER (2020011479). IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER (2020011479) FOR SIMILAR EVENT AND NO OTHER COMPLAINT WAS IDENTIFIED. BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR AND THE REPORTED EVENT WAS CONFIRMED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Additional Manufacturer Narrative · 1

(B)(4). PATIENT IDENTIFIER UNKNOWN / NOT PROVIDED. AGE AND DATE OF BIRTH UNKNOWN / NOT PROVIDED. PATIENT SEX UNKNOWN / NOT PROVIDED. WEIGHT UNKNOWN / NOT PROVIDED. EMAIL ADDRESS UNKNOWN / NOT PROVIDED. PMA/510(K) K011245 AND K002188.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT SURGERY THE MOUNT DAMAGED THE IMPLANT THREADS, IMPLANT WAS REMOVED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
742845 IMPL TAPERED SP 4.8MM 10M M OCTAGON DENTAL IMPLANT DZE ZIMMER DENTAL SPWB10 2020011479 00889024017566

Patients

Seq Age Sex Outcome Treatment
1