FDA Adverse Event Malfunction Summary report: N

COUPLER AC ZOOM

MDR report key: 10276089 · Received July 15, 2020

Report

Report Number
1221934-2020-01829
Event Type
Malfunction
Date Received
July 15, 2020
Date of Event
November 5, 2019
Report Date
July 9, 2020
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
FEM
UDI-DI
10886705028634
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. ADDITIONAL NARRATIVE: UDI: (B)(4). INVESTIGATION SUMMARY: THE DEVICE WAS RECEIVED AND EVALUATED AT THE SERVICE CENTER. THE REPORTED COMPLAINT THAT THE COUPLER AC ZOOM WAS BLURRY, WAS CONFIRMED. THE DEVICE WAS TESTED AND FOUND THAT THE DEVICE COULD NOT FOCUS DUE TO THE PRESENCE OF FOREIGN MATTER / PRECIPITATE. ALSO, THE GRASPING MECHANISM AND IMAGE DECENTERING WERE DAMAGED. FURTHER, THE ADJUSTING RINGS WERE EXTERNALLY DAMAGED. THE ENDOSCOPE GRASPING MECHANISM, ALONG WITH BOTH ADJUSTING RINGS INCLUDING O-RING SEALS WERE EXCHANGED, THE DEVICE WAS REPAIRED, TESTED AND FOUND TO BE WORKING ACCORDING TO SPECIFICATIONS. THE CAUSE OF THE IDENTIFIED FAILURES WERE IDENTIFIED BY THE SUPPLIER TO BE INCORRECT HANDLING OF THE DEVICE AND A POSSIBLE FALL. THE DEVICE MANUFACTURING HISTORY RECORD FOR CATALOG # 242436, SN : B000177 INDICATES 31 DEVICES WERE MANUFACTURED WITH THE WORK ORDER AND ACCEPTED INTO FINAL STOCK IN JAN 2018 WITH NO REPORTED DISCREPANCIES. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. DEPUY MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT DURING A SHOULDER ARTHROSCOPY, IT WAS OBSERVED THAT THE COUPLER AC ZOOM WAS BLURRY. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE SURGERY WITHOUT DELAY. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE WOULD NOT FOCUS DUE TO THE PRESENCE OF FOREIGN MATTER / PRECIPITATE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
742437 COUPLER AC ZOOM ENDOSCOPE CAMERA ADAPTOR FEM MEDOS INTERNATIONAL SARL 242436 10886705028634

Patients

Seq Age Sex Outcome Treatment
1