COUPLER AC ZOOM
Report
- Report Number
- 1221934-2020-01829
- Event Type
- Malfunction
- Date Received
- July 15, 2020
- Date of Event
- November 5, 2019
- Report Date
- July 9, 2020
- Manufacturer
- MEDOS INTERNATIONAL SARL
- Product Code
- FEM
- UDI-DI
- 10886705028634
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. ADDITIONAL NARRATIVE: UDI: (B)(4). INVESTIGATION SUMMARY: THE DEVICE WAS RECEIVED AND EVALUATED AT THE SERVICE CENTER. THE REPORTED COMPLAINT THAT THE COUPLER AC ZOOM WAS BLURRY, WAS CONFIRMED. THE DEVICE WAS TESTED AND FOUND THAT THE DEVICE COULD NOT FOCUS DUE TO THE PRESENCE OF FOREIGN MATTER / PRECIPITATE. ALSO, THE GRASPING MECHANISM AND IMAGE DECENTERING WERE DAMAGED. FURTHER, THE ADJUSTING RINGS WERE EXTERNALLY DAMAGED. THE ENDOSCOPE GRASPING MECHANISM, ALONG WITH BOTH ADJUSTING RINGS INCLUDING O-RING SEALS WERE EXCHANGED, THE DEVICE WAS REPAIRED, TESTED AND FOUND TO BE WORKING ACCORDING TO SPECIFICATIONS. THE CAUSE OF THE IDENTIFIED FAILURES WERE IDENTIFIED BY THE SUPPLIER TO BE INCORRECT HANDLING OF THE DEVICE AND A POSSIBLE FALL. THE DEVICE MANUFACTURING HISTORY RECORD FOR CATALOG # 242436, SN : B000177 INDICATES 31 DEVICES WERE MANUFACTURED WITH THE WORK ORDER AND ACCEPTED INTO FINAL STOCK IN JAN 2018 WITH NO REPORTED DISCREPANCIES. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. DEPUY MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.
IT WAS REPORTED BY THE CUSTOMER THAT DURING A SHOULDER ARTHROSCOPY, IT WAS OBSERVED THAT THE COUPLER AC ZOOM WAS BLURRY. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE SURGERY WITHOUT DELAY. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE WOULD NOT FOCUS DUE TO THE PRESENCE OF FOREIGN MATTER / PRECIPITATE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 742437 | COUPLER AC ZOOM | ENDOSCOPE CAMERA ADAPTOR | FEM | MEDOS INTERNATIONAL SARL | 242436 | 10886705028634 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |