FDA Adverse Event Malfunction Summary report: N

CHANNEL FLAT DRAIN WITH TROCAR

MDR report key: 10275751 · Received July 15, 2020

Report

Report Number
8030107-2020-00017
Event Type
Malfunction
Date Received
July 15, 2020
Report Date
July 15, 2020
Manufacturer
DEGANIA SILICONE LTD
Product Code
OTK
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

SINCE THERE WERE NO NON-CONFORMITIES DETECTED DURING BATCH HISTORY RECORDS, AND THE IN-PROCESS TEST RESULTS WERE WELL ABOVE THE REQUIRED MINIMUM, WE CONCLUDE THAT THERE WAS NO PROBLEM WITH OUR PART. THE DRAIN MOST LIKELY WAS DAMAGED IN USE AND TORE IN THE DAMAGED AREA; SILICONE DRAINS ARE KNOWN FOR THEIR ABILITY EASILY SNAP IF THEY WERE EVEN LIGHTLY CUT OR NICKED. THE COMPLAINT CONSIDERED TO BE NOT JUSTIFIED.

Description of Event or Problem · 1

BROKEN/CRACKED/FRACTURED DURING USE: FRACTURE AT THE DRAIN TUBE JOINT DURING USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
743164 CHANNEL FLAT DRAIN WITH TROCAR SILICONE FLUTED FLAT DRAIN 10 MM OTK DEGANIA SILICONE LTD 10 MM P1757997