FDA Adverse Event
Malfunction
Summary report: N
CHANNEL FLAT DRAIN WITH TROCAR
MDR report key: 10275751
·
Received July 15, 2020
Report
- Report Number
- 8030107-2020-00017
- Event Type
- Malfunction
- Date Received
- July 15, 2020
- Report Date
- July 15, 2020
- Manufacturer
- DEGANIA SILICONE LTD
- Product Code
- OTK
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- RISK MANAGER
Narratives
Additional Manufacturer Narrative · 1
SINCE THERE WERE NO NON-CONFORMITIES DETECTED DURING BATCH HISTORY RECORDS, AND THE IN-PROCESS TEST RESULTS WERE WELL ABOVE THE REQUIRED MINIMUM, WE CONCLUDE THAT THERE WAS NO PROBLEM WITH OUR PART. THE DRAIN MOST LIKELY WAS DAMAGED IN USE AND TORE IN THE DAMAGED AREA; SILICONE DRAINS ARE KNOWN FOR THEIR ABILITY EASILY SNAP IF THEY WERE EVEN LIGHTLY CUT OR NICKED. THE COMPLAINT CONSIDERED TO BE NOT JUSTIFIED.
Description of Event or Problem · 1
BROKEN/CRACKED/FRACTURED DURING USE: FRACTURE AT THE DRAIN TUBE JOINT DURING USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 743164 | CHANNEL FLAT DRAIN WITH TROCAR | SILICONE FLUTED FLAT DRAIN 10 MM | OTK | DEGANIA SILICONE LTD | 10 MM | P1757997 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |