OMNICELL I.V.STATION ONCO
Report
- Report Number
- 3011278888-2020-00012
- Event Type
- Malfunction
- Date Received
- July 14, 2020
- Date of Event
- June 30, 2020
- Report Date
- May 20, 2021
- Manufacturer
- OMNICELL, INC.
- Product Code
- NEP
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- 003
Narratives
AS OF (B)(6) 2020, THE ESTABLISHMENT REGISTRATION AND LISTING FOR THIS DEVICE WAS UPDATED TO OMNICELL, INC. FROM AESYNT, INC, WHICH WAS NOT REFLECTED IN THE ORIGINAL REPORT SUBMISSION. THEREFORE, THIS REPORT IS A CORRECTION TO THE MANUFACTURER LISTED IN SECTIONS D3 AND G1.
ON (B)(6) 2020 A SPILL INVOLVING THE CHEMOTHERAPY DRUG PREPARATION PACLITAXEL OCCURRED INSIDE THE I.V. STATION ONCO DEVICE. THE SPILL WAS IDENTIFIED AND SUBSEQUENTLY CLEANED BY THE USER. THE DRUG WAS SEEPING FROM THE LOADING SYRINGE CAUSING DRIPS TO ACCUMULATE ON THE BAG CLAMPS. MODIFICATIONS WERE MADE TO DRUG AND SYRINGE PRESSURE, VOLUME, SPEED AND TIME PARAMETERS TO PREVENT RECURRENCE OF THE DRIPPING. THERE IS NO ADVERSE PATIENT EFFECT RELATED TO THIS DRUG SPILL WITHIN THE DEVICE.