ALL-IN-ONE CCU+LIGHT ROW
Report
- Report Number
- 1221934-2020-01819
- Event Type
- Malfunction
- Date Received
- July 14, 2020
- Date of Event
- January 1, 2018
- Report Date
- June 16, 2020
- Manufacturer
- NO.100 NAN GU ROAD
- Product Code
- FCW
- UDI-DI
- 10886705028702
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- OTHER
Narratives
UDI: (B)(4). INVESTIGATION SUMMARY: THE COMPLAINT DEVICE WAS RECEIVED AT THE SERVICE CENTER AND EVALUATED. IT WAS REPORTED THAT THE DEVICE DID NOT DISPLAY THE SCREEN. PER SERVICE MANUAL OPERATIONAL & DIAGNOSTIC AND SERVICE REPORTS, THIS COMPLAINT CAN BE CONFIRMED. FOLLOWING DEFECTS WERE IDENTIFIED UPON EVALUATION OF THE COMPLAINT DEVICE: ACCESSORY DEVICE WAS NOT RECOGNIZED, INTERMITTENT OPERATION, NO IMAGE WAS SHOWN ON THE DISPLAY, SYSTEM SOFTWARE WAS DEFECTIVE, MEMORY BACKUP BATTERY WAS DAMAGED, CARRIER BOARD, FPGA BOARD, FRONT BOARD AND QSEVEN MODULE WERE DEFECTIVE, SOCKET, HARD DRIVE AND CONNECTING CABLE WERE DEFECTIVE. ALL THE DEFECTIVE PARTS WERE REPLACED AND THE SOFTWARE WAS RE-INSTALLED TO RESOLVE THE IDENTIFIED FAILURES. WITH THE AVAILABLE INFORMATION, WE CANNOT DETERMINE THE ROOT CAUSE OF THE IDENTIFIED FAILURES. THE FAULTY COMPONENTS OF THE DEVICE AND DEFECTIVE SOFTWARE WOULD HAVE CAUSED THE CUSTOMER TO EXPERIENCE THE REPORTED PROBLEM. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE FINISHED DEVICE SERIAL NUMBER (B)(4), AND THE CERTIFICATE OF CONFORMANCE STATED THAT THE DEVICE WAS MANUFACTURED IN COMPLIANCE WITH THE QUALITY SYSTEM RELEASE CRITERIA. SERVICE AND REPAIR FUNCTIONS WERE PERFORMED TO ADDRESS NCR (B)(4). IT WAS CARRIED OUT TO RESOLVE THE DISCREPANCY OF THREE OF THE USB PORTS ON THE BACK OF THE DEVICE NOT DISPLAYING THE KEYBOARD WHEN PLUGGED INTO. BY ADDRESSING NCR (B)(4), IT WAS VERIFIED THAT THE KEYBOARD ICON APPEARS IN THE UPPER-RIGHT CORNER OF THE PURVUE LCD SCREEN WHEN A USB KEYBOARD IS CONNECTED TO EACH OF THE NINE USB PORTS. ALSO, SERVICE AND REPAIR FUNCTIONS WERE PERFORMED TO ADDRESS NCR (B)(4). IT WAS CARRIED OUT TO ENSURE THAT THE LIVE VIDEO APPEARS PROPERLY ON THE NDS SURGICAL DISPLAY WITH THE PURVUE CAMERA HEAD AND LIGHT CABLE CONNECTED. AT THIS POINT IN TIME, NO OTHER CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT PREOPERATIVELY TO AN UNSPECIFIED SURGERY, IT WAS OBSERVED THAT THE ALL IN ONE CCU+LIGHT ROW DEVICE DID NOT DISPLAY THE SCREEN. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD INTERMITTENT OPERATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 735687 | ALL-IN-ONE CCU+LIGHT ROW | ENDOSCOPIC VIDEO IMAGING SYSTEM | FCW | NO.100 NAN GU ROAD | 242302 | 10886705028702 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |