FDA Adverse Event Malfunction Summary report: N

ALL-IN-ONE CCU+LIGHT ROW

MDR report key: 10271461 · Received July 14, 2020

Report

Report Number
1221934-2020-01819
Event Type
Malfunction
Date Received
July 14, 2020
Date of Event
January 1, 2018
Report Date
June 16, 2020
Manufacturer
NO.100 NAN GU ROAD
Product Code
FCW
UDI-DI
10886705028702
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UDI: (B)(4). INVESTIGATION SUMMARY: THE COMPLAINT DEVICE WAS RECEIVED AT THE SERVICE CENTER AND EVALUATED. IT WAS REPORTED THAT THE DEVICE DID NOT DISPLAY THE SCREEN. PER SERVICE MANUAL OPERATIONAL & DIAGNOSTIC AND SERVICE REPORTS, THIS COMPLAINT CAN BE CONFIRMED. FOLLOWING DEFECTS WERE IDENTIFIED UPON EVALUATION OF THE COMPLAINT DEVICE: ACCESSORY DEVICE WAS NOT RECOGNIZED, INTERMITTENT OPERATION, NO IMAGE WAS SHOWN ON THE DISPLAY, SYSTEM SOFTWARE WAS DEFECTIVE, MEMORY BACKUP BATTERY WAS DAMAGED, CARRIER BOARD, FPGA BOARD, FRONT BOARD AND QSEVEN MODULE WERE DEFECTIVE, SOCKET, HARD DRIVE AND CONNECTING CABLE WERE DEFECTIVE. ALL THE DEFECTIVE PARTS WERE REPLACED AND THE SOFTWARE WAS RE-INSTALLED TO RESOLVE THE IDENTIFIED FAILURES. WITH THE AVAILABLE INFORMATION, WE CANNOT DETERMINE THE ROOT CAUSE OF THE IDENTIFIED FAILURES. THE FAULTY COMPONENTS OF THE DEVICE AND DEFECTIVE SOFTWARE WOULD HAVE CAUSED THE CUSTOMER TO EXPERIENCE THE REPORTED PROBLEM. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE FINISHED DEVICE SERIAL NUMBER (B)(4), AND THE CERTIFICATE OF CONFORMANCE STATED THAT THE DEVICE WAS MANUFACTURED IN COMPLIANCE WITH THE QUALITY SYSTEM RELEASE CRITERIA. SERVICE AND REPAIR FUNCTIONS WERE PERFORMED TO ADDRESS NCR (B)(4). IT WAS CARRIED OUT TO RESOLVE THE DISCREPANCY OF THREE OF THE USB PORTS ON THE BACK OF THE DEVICE NOT DISPLAYING THE KEYBOARD WHEN PLUGGED INTO. BY ADDRESSING NCR (B)(4), IT WAS VERIFIED THAT THE KEYBOARD ICON APPEARS IN THE UPPER-RIGHT CORNER OF THE PURVUE LCD SCREEN WHEN A USB KEYBOARD IS CONNECTED TO EACH OF THE NINE USB PORTS. ALSO, SERVICE AND REPAIR FUNCTIONS WERE PERFORMED TO ADDRESS NCR (B)(4). IT WAS CARRIED OUT TO ENSURE THAT THE LIVE VIDEO APPEARS PROPERLY ON THE NDS SURGICAL DISPLAY WITH THE PURVUE CAMERA HEAD AND LIGHT CABLE CONNECTED. AT THIS POINT IN TIME, NO OTHER CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.

Description of Event or Problem · 1

IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT PREOPERATIVELY TO AN UNSPECIFIED SURGERY, IT WAS OBSERVED THAT THE ALL IN ONE CCU+LIGHT ROW DEVICE DID NOT DISPLAY THE SCREEN. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD INTERMITTENT OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
735687 ALL-IN-ONE CCU+LIGHT ROW ENDOSCOPIC VIDEO IMAGING SYSTEM FCW NO.100 NAN GU ROAD 242302 10886705028702

Patients

Seq Age Sex Outcome Treatment
1