FDA Adverse Event Malfunction Summary report: N

BD INSULIN SYRINGE WITH THE BD ULTRA-FINE NEEDLE

MDR report key: 10269456 · Received July 14, 2020

Report

Report Number
1920898-2020-00843
Event Type
Malfunction
Date Received
July 14, 2020
Date of Event
June 23, 2020
Report Date
August 12, 2020
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
UDI-DI
00382908431033
PMA / PMN Number
K024112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES D.10. RETURNED TO MANUFACTURER ON: 7/29/2020 H.6. INVESTIGATION: CUSTOMER RETURNED ONE (1) 0.3ML BD INSULIN SYRINGE FROM LOT 9231323. CONSUMER REPORTED NEEDLE SHIELD DIFFICULT TO REMOVE. NEEDLE HUB SEPARATES AND STAYS IN THE SHIELD. THE RETURNED SYRINGE WAS EXAMINED, AND IT WAS OBSERVED THAT THE NEEDLE HUB/SHIELD ASSEMBLY WAS SEPARATED FROM THE SYRINGE BARREL; NO DAMAGE TO THE BARREL TIP WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 9231323. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. PROCESS SUMMARY: AUTOMATIC SYRINGE ASSEMBLY MACHINE, WHICH FEEDS 3/10CC, SYRINGE COMPONENTS (BARREL, STOPPER, PLUNGER, NEEDLE ASSEMBLY & CAP) AND ASSEMBLES THESE COMPONENTS. THIS MACHINE CONSISTS OF A BARREL CLEANING DIAL, LUBRICATION DIAL, PLUNGER/STOPPER ASSEMBLY DIAL, SYRINGE ASSEMBLY DIAL, AND VARIOUS INSPECTIONS AND TRANSFER DIALS. DHR, L2L DISPATCHES, AND LOGBOOK ENTRIES WERE LOOKED AT, NOTHING PERTAINING TO THIS DEFECT WAS FOUND. ROOT CAUSE FOR THIS DEFECT CANNOT BE DETERMINED. CAPA#1630423 WAS INITIATED. SEE H.10.

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED AS OF (B)(6)2020 THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 9231323. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD ULTRA-FINE¿ NEEDLE HUB SEPARATED FROM THE BD INSULIN SYRINGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "NEEDLE SHIELD DIFFICULT TO REMOVE. NEEDLE HUB SEPARATES AND STAYS IN THE SHIELD."

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD ULTRA-FINE¿ NEEDLE HUB SEPARATED FROM THE BD INSULIN SYRINGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "NEEDLE SHIELD DIFFICULT TO REMOVE. NEEDLE HUB SEPARATES AND STAYS IN THE SHIELD.".

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BD ULTRA-FINE¿ NEEDLE HUB SEPARATED FROM THE BD INSULIN SYRINGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "NEEDLE SHIELD DIFFICULT TO REMOVE. NEEDLE HUB SEPARATES AND STAYS IN THE SHIELD."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
733041 BD INSULIN SYRINGE WITH THE BD ULTRA-FINE NEEDLE PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 328431 9231323 00382908431033

Patients

Seq Age Sex Outcome Treatment
1 Other