BD INSULIN SYRINGE WITH THE BD ULTRA-FINE NEEDLE
Report
- Report Number
- 1920898-2020-00843
- Event Type
- Malfunction
- Date Received
- July 14, 2020
- Date of Event
- June 23, 2020
- Report Date
- August 12, 2020
- Manufacturer
- BD MEDICAL - DIABETES CARE
- Product Code
- FMF
- UDI-DI
- 00382908431033
- PMA / PMN Number
- K024112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- 003
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D.10. DEVICE AVAILABLE FOR EVAL?: YES D.10. RETURNED TO MANUFACTURER ON: 7/29/2020 H.6. INVESTIGATION: CUSTOMER RETURNED ONE (1) 0.3ML BD INSULIN SYRINGE FROM LOT 9231323. CONSUMER REPORTED NEEDLE SHIELD DIFFICULT TO REMOVE. NEEDLE HUB SEPARATES AND STAYS IN THE SHIELD. THE RETURNED SYRINGE WAS EXAMINED, AND IT WAS OBSERVED THAT THE NEEDLE HUB/SHIELD ASSEMBLY WAS SEPARATED FROM THE SYRINGE BARREL; NO DAMAGE TO THE BARREL TIP WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 9231323. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. PROCESS SUMMARY: AUTOMATIC SYRINGE ASSEMBLY MACHINE, WHICH FEEDS 3/10CC, SYRINGE COMPONENTS (BARREL, STOPPER, PLUNGER, NEEDLE ASSEMBLY & CAP) AND ASSEMBLES THESE COMPONENTS. THIS MACHINE CONSISTS OF A BARREL CLEANING DIAL, LUBRICATION DIAL, PLUNGER/STOPPER ASSEMBLY DIAL, SYRINGE ASSEMBLY DIAL, AND VARIOUS INSPECTIONS AND TRANSFER DIALS. DHR, L2L DISPATCHES, AND LOGBOOK ENTRIES WERE LOOKED AT, NOTHING PERTAINING TO THIS DEFECT WAS FOUND. ROOT CAUSE FOR THIS DEFECT CANNOT BE DETERMINED. CAPA#1630423 WAS INITIATED. SEE H.10.
H.6. INVESTIGATION: NO SAMPLES (INCLUDING PHOTOS) WERE RETURNED AS OF (B)(6)2020 THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED AND THE ROOT CAUSE IS UNDETERMINED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 9231323. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO (0) NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT.
IT WAS REPORTED THAT THE BD ULTRA-FINE¿ NEEDLE HUB SEPARATED FROM THE BD INSULIN SYRINGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "NEEDLE SHIELD DIFFICULT TO REMOVE. NEEDLE HUB SEPARATES AND STAYS IN THE SHIELD."
IT WAS REPORTED THAT THE BD ULTRA-FINE¿ NEEDLE HUB SEPARATED FROM THE BD INSULIN SYRINGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "NEEDLE SHIELD DIFFICULT TO REMOVE. NEEDLE HUB SEPARATES AND STAYS IN THE SHIELD.".
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED. (B)(4).
IT WAS REPORTED THAT THE BD ULTRA-FINE¿ NEEDLE HUB SEPARATED FROM THE BD INSULIN SYRINGE DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: "NEEDLE SHIELD DIFFICULT TO REMOVE. NEEDLE HUB SEPARATES AND STAYS IN THE SHIELD."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 733041 | BD INSULIN SYRINGE WITH THE BD ULTRA-FINE NEEDLE | PISTON SYRINGE | FMF | BD MEDICAL - DIABETES CARE | 328431 | 9231323 | 00382908431033 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |